Abstract
We propose and investigate weighted expectation maximization (EM) algorithms for image reconstruction in x-ray tomography. The development of the algorithms is motivated by the respiratory-gated megavoltage tomography problem, in which the acquired asymmetric cone-beam projections are limited in number and unevenly sampled over view angle. In these cases, images reconstructed by use of the conventional EM algorithm can contain ring- and streak-like artefacts that are attributable to a combination of data inconsistencies and truncation of the projection data. By use of computer-simulated and clinical gated fan-beam megavoltage projection data, we demonstrate that the proposed weighted EM algorithms effectively mitigate such image artefacts.
| Original language | English (US) |
|---|---|
| Pages (from-to) | N299-N307 |
| Journal | Physics in Medicine and Biology |
| Volume | 50 |
| Issue number | 21 |
| DOIs | |
| State | Published - Nov 7 2005 |
| Externally published | Yes |