Unexpected loss of bipolar lead function associated with lead connection design in dual chamber pacemakers

Charles C. Gornick, Adrian Almquist, Linda Kallinen, Margaret Fashingbauer, Ann Dunnigan, Chuen Tang, Robert G. Hauser, Simon Milstein

Research output: Contribution to journalArticlepeer-review

1 Scopus citations

Abstract

Unexpected loss of bipolar pacing function can occur in Guidant VIGOR dual chamber pacemakers with QC2 (Quick Connect) headers. In a retrospective review of 305 implanted patients, 23 exhibited unexpected loss of bipolar pacing lead function from < 1 to 51 months postimplant. Atrial lead problems were seen in 17 patients and ventricular lead problems in 6 (P < 0.05 atrial vs ventricular). Significant symptoms developed in six patients including two with syncope as a result of noncapture. No significant change in bipolar sensing function was seen. Reprogramming devices to the unipolar mode restored appropriate pacing thresholds and lead impedance. The changes in bipolar lead function are likely caused by failure of the QC2 header lead connection system over time.

Original languageEnglish (US)
Pages (from-to)1394-1396
Number of pages3
JournalPACE - Pacing and Clinical Electrophysiology
Volume25
Issue number9
DOIs
StatePublished - Sep 1 2002

Keywords

  • Equipment failure
  • Pacemaker
  • Syncope

Fingerprint

Dive into the research topics of 'Unexpected loss of bipolar lead function associated with lead connection design in dual chamber pacemakers'. Together they form a unique fingerprint.