The Proview phosphene tonometer fails to measure ocular pressure accurately in clinical practice

Tara L. Alvarez, Stephen A. Gollance, Gordon A. Thomas, Richard J. Greene, Peter M. Marchetto, Eugene J. Moore, Tony Realini, Jeffrey M. Liebmann, Robert Ritch, Paul Lama, Robert D. Fechtner

Research output: Contribution to journalArticle

36 Citations (Scopus)

Abstract

Purpose To evaluate the Proview Eye Pressure Monitor as a medical instrument and as a technique for enabling a patient to obtain an accurate measure of his or her intraocular pressure (IOP). Design An experimental laboratory evaluation and an independent prospective clinical study to test the reproducibility and accuracy of the Proview technique relative to Goldmann applanation tonometry. Participants For the laboratory study, we analyzed 3 tonometers, each packaged as a Proview Eye Pressure Monitor by Bausch & Lomb. In the independent prospective experimental study, 137 subjects participated, consisting of healthy volunteers and glaucoma patients. Methods For laboratory testing, we held each tonometer with a micrometer to assure controlled positioning and pressed its sensing tip against a force meter that produced a calibrated, digital force reading. For clinical testing, we taught subjects (n = 137) to use the Proview technique in accordance with the manufacturer's instructions. Each subject obtained 5 measurements with each of the 5 different Proview devices. A clinician measured the IOP using Goldmann applanation tonometry. Main outcome measures We measured the absolute value, linearity, and repeatability of the force meter readings on the tonometers during the instrument laboratory evaluation. The accuracy was evaluated by comparing the Proview measurements to the Goldmann applanation measurements. Reproducibility of clinical Proview measurements was also measured. All measurements were in mmHg during the clinical evaluation. Results Laboratory: There was a linear relationship between the pressures read by the Proview tonometers and known forces. The Proview tonometers read the maximum pressure applied. Clinical: The Proview technique is simple to use because it was comfortable and reproducible, with an average variance of the measurements by the same patient of 3.4 mmHg2. Other variables besides IOP seem to affect the Proview pressure measurements, as seen in the large scatter in our data, measured by our correlation coefficient of r = 0.41. The sensitivity of the Proview technique to detect patients with high IOP (which we defined as a Goldmann pressure of ≥22 mmHg) is low; the Proview pressure identified only 18% (4/22) of these patients. Conclusions The Proview instrument and technique were reproducible. However, the Proview tonometer seems not to be reliable as an indicator of IOP. The sensitivity for detecting high IOP was low in this cohort, and the agreement with Goldmann applanation was poor for some individuals. This brings into question the underlying assumption that a force proportional to the IOP generates phosphenes.

Original languageEnglish (US)
Pages (from-to)1077-1085
Number of pages9
JournalOphthalmology
Volume111
Issue number6
DOIs
StatePublished - Jun 1 2004

Fingerprint

Phosphenes
Intraocular Pressure
Pressure
Manometry
Reading
Prospective Studies
Glaucoma
Healthy Volunteers
Research Design
Outcome Assessment (Health Care)
Equipment and Supplies

Cite this

Alvarez, T. L., Gollance, S. A., Thomas, G. A., Greene, R. J., Marchetto, P. M., Moore, E. J., ... Fechtner, R. D. (2004). The Proview phosphene tonometer fails to measure ocular pressure accurately in clinical practice. Ophthalmology, 111(6), 1077-1085. https://doi.org/10.1016/j.ophtha.2003.09.036

The Proview phosphene tonometer fails to measure ocular pressure accurately in clinical practice. / Alvarez, Tara L.; Gollance, Stephen A.; Thomas, Gordon A.; Greene, Richard J.; Marchetto, Peter M.; Moore, Eugene J.; Realini, Tony; Liebmann, Jeffrey M.; Ritch, Robert; Lama, Paul; Fechtner, Robert D.

In: Ophthalmology, Vol. 111, No. 6, 01.06.2004, p. 1077-1085.

Research output: Contribution to journalArticle

Alvarez, TL, Gollance, SA, Thomas, GA, Greene, RJ, Marchetto, PM, Moore, EJ, Realini, T, Liebmann, JM, Ritch, R, Lama, P & Fechtner, RD 2004, 'The Proview phosphene tonometer fails to measure ocular pressure accurately in clinical practice', Ophthalmology, vol. 111, no. 6, pp. 1077-1085. https://doi.org/10.1016/j.ophtha.2003.09.036
Alvarez, Tara L. ; Gollance, Stephen A. ; Thomas, Gordon A. ; Greene, Richard J. ; Marchetto, Peter M. ; Moore, Eugene J. ; Realini, Tony ; Liebmann, Jeffrey M. ; Ritch, Robert ; Lama, Paul ; Fechtner, Robert D. / The Proview phosphene tonometer fails to measure ocular pressure accurately in clinical practice. In: Ophthalmology. 2004 ; Vol. 111, No. 6. pp. 1077-1085.
@article{0ac3571e751a44f8b146c9a5f26f0f43,
title = "The Proview phosphene tonometer fails to measure ocular pressure accurately in clinical practice",
abstract = "Purpose To evaluate the Proview Eye Pressure Monitor as a medical instrument and as a technique for enabling a patient to obtain an accurate measure of his or her intraocular pressure (IOP). Design An experimental laboratory evaluation and an independent prospective clinical study to test the reproducibility and accuracy of the Proview technique relative to Goldmann applanation tonometry. Participants For the laboratory study, we analyzed 3 tonometers, each packaged as a Proview Eye Pressure Monitor by Bausch & Lomb. In the independent prospective experimental study, 137 subjects participated, consisting of healthy volunteers and glaucoma patients. Methods For laboratory testing, we held each tonometer with a micrometer to assure controlled positioning and pressed its sensing tip against a force meter that produced a calibrated, digital force reading. For clinical testing, we taught subjects (n = 137) to use the Proview technique in accordance with the manufacturer's instructions. Each subject obtained 5 measurements with each of the 5 different Proview devices. A clinician measured the IOP using Goldmann applanation tonometry. Main outcome measures We measured the absolute value, linearity, and repeatability of the force meter readings on the tonometers during the instrument laboratory evaluation. The accuracy was evaluated by comparing the Proview measurements to the Goldmann applanation measurements. Reproducibility of clinical Proview measurements was also measured. All measurements were in mmHg during the clinical evaluation. Results Laboratory: There was a linear relationship between the pressures read by the Proview tonometers and known forces. The Proview tonometers read the maximum pressure applied. Clinical: The Proview technique is simple to use because it was comfortable and reproducible, with an average variance of the measurements by the same patient of 3.4 mmHg2. Other variables besides IOP seem to affect the Proview pressure measurements, as seen in the large scatter in our data, measured by our correlation coefficient of r = 0.41. The sensitivity of the Proview technique to detect patients with high IOP (which we defined as a Goldmann pressure of ≥22 mmHg) is low; the Proview pressure identified only 18{\%} (4/22) of these patients. Conclusions The Proview instrument and technique were reproducible. However, the Proview tonometer seems not to be reliable as an indicator of IOP. The sensitivity for detecting high IOP was low in this cohort, and the agreement with Goldmann applanation was poor for some individuals. This brings into question the underlying assumption that a force proportional to the IOP generates phosphenes.",
author = "Alvarez, {Tara L.} and Gollance, {Stephen A.} and Thomas, {Gordon A.} and Greene, {Richard J.} and Marchetto, {Peter M.} and Moore, {Eugene J.} and Tony Realini and Liebmann, {Jeffrey M.} and Robert Ritch and Paul Lama and Fechtner, {Robert D.}",
year = "2004",
month = "6",
day = "1",
doi = "10.1016/j.ophtha.2003.09.036",
language = "English (US)",
volume = "111",
pages = "1077--1085",
journal = "Ophthalmology",
issn = "0161-6420",
publisher = "Elsevier Inc.",
number = "6",

}

TY - JOUR

T1 - The Proview phosphene tonometer fails to measure ocular pressure accurately in clinical practice

AU - Alvarez, Tara L.

AU - Gollance, Stephen A.

AU - Thomas, Gordon A.

AU - Greene, Richard J.

AU - Marchetto, Peter M.

AU - Moore, Eugene J.

AU - Realini, Tony

AU - Liebmann, Jeffrey M.

AU - Ritch, Robert

AU - Lama, Paul

AU - Fechtner, Robert D.

PY - 2004/6/1

Y1 - 2004/6/1

N2 - Purpose To evaluate the Proview Eye Pressure Monitor as a medical instrument and as a technique for enabling a patient to obtain an accurate measure of his or her intraocular pressure (IOP). Design An experimental laboratory evaluation and an independent prospective clinical study to test the reproducibility and accuracy of the Proview technique relative to Goldmann applanation tonometry. Participants For the laboratory study, we analyzed 3 tonometers, each packaged as a Proview Eye Pressure Monitor by Bausch & Lomb. In the independent prospective experimental study, 137 subjects participated, consisting of healthy volunteers and glaucoma patients. Methods For laboratory testing, we held each tonometer with a micrometer to assure controlled positioning and pressed its sensing tip against a force meter that produced a calibrated, digital force reading. For clinical testing, we taught subjects (n = 137) to use the Proview technique in accordance with the manufacturer's instructions. Each subject obtained 5 measurements with each of the 5 different Proview devices. A clinician measured the IOP using Goldmann applanation tonometry. Main outcome measures We measured the absolute value, linearity, and repeatability of the force meter readings on the tonometers during the instrument laboratory evaluation. The accuracy was evaluated by comparing the Proview measurements to the Goldmann applanation measurements. Reproducibility of clinical Proview measurements was also measured. All measurements were in mmHg during the clinical evaluation. Results Laboratory: There was a linear relationship between the pressures read by the Proview tonometers and known forces. The Proview tonometers read the maximum pressure applied. Clinical: The Proview technique is simple to use because it was comfortable and reproducible, with an average variance of the measurements by the same patient of 3.4 mmHg2. Other variables besides IOP seem to affect the Proview pressure measurements, as seen in the large scatter in our data, measured by our correlation coefficient of r = 0.41. The sensitivity of the Proview technique to detect patients with high IOP (which we defined as a Goldmann pressure of ≥22 mmHg) is low; the Proview pressure identified only 18% (4/22) of these patients. Conclusions The Proview instrument and technique were reproducible. However, the Proview tonometer seems not to be reliable as an indicator of IOP. The sensitivity for detecting high IOP was low in this cohort, and the agreement with Goldmann applanation was poor for some individuals. This brings into question the underlying assumption that a force proportional to the IOP generates phosphenes.

AB - Purpose To evaluate the Proview Eye Pressure Monitor as a medical instrument and as a technique for enabling a patient to obtain an accurate measure of his or her intraocular pressure (IOP). Design An experimental laboratory evaluation and an independent prospective clinical study to test the reproducibility and accuracy of the Proview technique relative to Goldmann applanation tonometry. Participants For the laboratory study, we analyzed 3 tonometers, each packaged as a Proview Eye Pressure Monitor by Bausch & Lomb. In the independent prospective experimental study, 137 subjects participated, consisting of healthy volunteers and glaucoma patients. Methods For laboratory testing, we held each tonometer with a micrometer to assure controlled positioning and pressed its sensing tip against a force meter that produced a calibrated, digital force reading. For clinical testing, we taught subjects (n = 137) to use the Proview technique in accordance with the manufacturer's instructions. Each subject obtained 5 measurements with each of the 5 different Proview devices. A clinician measured the IOP using Goldmann applanation tonometry. Main outcome measures We measured the absolute value, linearity, and repeatability of the force meter readings on the tonometers during the instrument laboratory evaluation. The accuracy was evaluated by comparing the Proview measurements to the Goldmann applanation measurements. Reproducibility of clinical Proview measurements was also measured. All measurements were in mmHg during the clinical evaluation. Results Laboratory: There was a linear relationship between the pressures read by the Proview tonometers and known forces. The Proview tonometers read the maximum pressure applied. Clinical: The Proview technique is simple to use because it was comfortable and reproducible, with an average variance of the measurements by the same patient of 3.4 mmHg2. Other variables besides IOP seem to affect the Proview pressure measurements, as seen in the large scatter in our data, measured by our correlation coefficient of r = 0.41. The sensitivity of the Proview technique to detect patients with high IOP (which we defined as a Goldmann pressure of ≥22 mmHg) is low; the Proview pressure identified only 18% (4/22) of these patients. Conclusions The Proview instrument and technique were reproducible. However, the Proview tonometer seems not to be reliable as an indicator of IOP. The sensitivity for detecting high IOP was low in this cohort, and the agreement with Goldmann applanation was poor for some individuals. This brings into question the underlying assumption that a force proportional to the IOP generates phosphenes.

UR - http://www.scopus.com/inward/record.url?scp=2942601243&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=2942601243&partnerID=8YFLogxK

U2 - 10.1016/j.ophtha.2003.09.036

DO - 10.1016/j.ophtha.2003.09.036

M3 - Article

C2 - 15177956

AN - SCOPUS:2942601243

VL - 111

SP - 1077

EP - 1085

JO - Ophthalmology

JF - Ophthalmology

SN - 0161-6420

IS - 6

ER -