The effect of a genetic variant on quantitative real-time PCR in a case of disseminated adenovirus infection

  • Thomas J. Gniadek
  • , Michael T. Forman
  • , Isabella Martin
  • , Ravit Arav-Boger
  • , Alexandra Valsamakis

Research output: Contribution to journalArticlepeer-review

Abstract

A patient developed disseminated adenovirus infection following bone marrow transplant. TaqMan real-time PCR showed reduced maximum fluorescence in the amplification curve from all plasma samples. Sequencing revealed three single nucleotide mismatches between the TaqMan probe and probe binding region. Real-time PCR with probe matching the isolate sequence showed normal amplification and a higher copy number result.

Original languageEnglish (US)
Pages (from-to)40-43
Number of pages4
JournalDiagnostic Microbiology and Infectious Disease
Volume89
Issue number1
DOIs
StatePublished - Sep 2017
Externally publishedYes

Bibliographical note

Publisher Copyright:
© 2017 Elsevier Inc.

UN SDGs

This output contributes to the following UN Sustainable Development Goals (SDGs)

  1. SDG 3 - Good Health and Well-being
    SDG 3 Good Health and Well-being

Keywords

  • Aberrant real-time PCR
  • Adenovirus
  • Reduced fluorescence