The arterial distribution of Embozene and Embosphere microspheres in sheep kidney and uterus embolization models

Valentin Verret, Sada Homayra Ghegediban, Michel Wassef, Jean Pierre Pelage, Jafar Golzarian, Alexandre Laurent

Research output: Contribution to journalArticlepeer-review

39 Scopus citations

Abstract

Purpose To compare the in vivo distribution of the new embolic Embozene versus Embosphere as a control in the sheep renal and uterine vasculature. Materials and Methods Twelve sheep (three per group of product and size) were selectively embolized with Embozene 700 μm, Embozene 900 μm, Embosphere 500-700 μm, or Embosphere 700-900μm, in one renal artery (0.5 mL microspheres) and in the two uterine arteries (0.25 mL each) and sacrificed 72 hours later for pathologic examination of kidney and uterus. Partition of microspheres in the vasculature was determined according to a classification of the renal and the uterine vasculatures into several zones. Vascular diameters and microsphere deformation were measured. Results Embozene 700 μm and Embozene 900 μm occluded significantly more distally than Embosphere 500-700μm and Embosphere 700-900 μm in renal and uterine vasculature. For Embozene, the vessel diameter was not significantly different between the two sizes, for each organ, whereas it was significantly larger for Embosphere 700-900 μm than for Embosphere 500-700 μm in each organ. Embozene deformation was significantly higher than that of Embosphere in renal and uterine vasculature, increased from proximal to distal in location for both organs and correlated negatively with vessel diameter (Rho = 0.623; P < .0001). Embosphere deformation did not vary according to the zone. Conclusions Embozene microspheres have a higher in vivo deformation, which results in a more distal occlusion within the vascular network compared with reference Embosphere microspheres. The diameter of occluded vessels varied for the tested size range for Embosphere but was independent of the tested microsphere size range used for Embozene. The deformation of Embozene appears to determine the size of the vessels occluded as opposed to the granulometric particle size, which makes level of occlusion unpredictable.

Original languageEnglish (US)
Pages (from-to)220-228
Number of pages9
JournalJournal of Vascular and Interventional Radiology
Volume22
Issue number2
DOIs
StatePublished - Feb 2011

Fingerprint

Dive into the research topics of 'The arterial distribution of Embozene and Embosphere microspheres in sheep kidney and uterus embolization models'. Together they form a unique fingerprint.

Cite this