Abstract
Objective. In photon counting detectors (PCDs), electric pulses induced by two or more x-ray photons can pile up and result in count losses when their temporal separation is less than the detector dead time. The correction of pulse pile-up-induced count loss is particularly difficult for paralyzable PCDs since a given value of recorded counts can correspond to two different values of true photon interactions. In contrast, charge (energy) integrating detectors work by integrating collected electric charge induced by x-rays over time and do not suffer from pile-up losses. This work introduces an inexpensive readout circuit element to the circuits of PCDs to simultaneously collect time-integrated charge to correct pile-up-induced count losses. Approach. Prototype electronics were constructed to collect time-integrated charges simultaneously with photon counts. A splitter was used to feed the electric signal in parallel to both a digital counter and a charge integrator. After recording PCD counts and integrating collected charge, a lookup table can be generated to map raw counts in the total- and high-energy bins and total charge to estimate pile-up-free true counts. Proof-of-concept imaging experiments were performed with a CdTe-based PCD array to test this method. Main results. The proposed electronics successfully recorded photon counts and time-integrated charge simultaneously, and whereas photon counts exhibited paralyzable pulse pile-up, time-integrated charge using the same electric signal as the counts measurement was linear with x-ray flux. With the proposed correction, paralyzable PCD counts became linear with input flux for both total- and high-energy bins. At high flux levels, uncorrected post-log measurements of PMMA objects severely overestimated radiological path lengths for both energy bins. After the proposed correction, the non-monotonic measurements again became linear with flux and accurately represented the true radiological path lengths. No impact on the spatial resolution was observed after the proposed correction in images of a line-pair test pattern. Significance. Time-integrated charge can be used to correct for pulse pile-up in paralyzable PCDs where analytical solutions may be difficult to use, and integrated charge can be collected simultaneously with counts using inexpensive electronics.
| Original language | English (US) |
|---|---|
| Article number | 155003 |
| Journal | Physics in Medicine and Biology |
| Volume | 68 |
| Issue number | 15 |
| DOIs | |
| State | Published - Aug 7 2023 |
| Externally published | Yes |
Bibliographical note
Publisher Copyright:© 2023 The Author(s). Published on behalf of Institute of Physics and Engineering in Medicine by IOP Publishing Ltd.
Keywords
- photon counting CT
- photon counting detector
- x-ray detector
- x-ray imaging
Fingerprint
Dive into the research topics of 'Simultaneous photon counting and charge integrating for pulse pile-up correction in paralyzable photon counting detectors'. Together they form a unique fingerprint.Cite this
- APA
- Standard
- Harvard
- Vancouver
- Author
- BIBTEX
- RIS