T1 - Safety and performance of a novel subxiphoidal pacemaker system
AU - Libbus, Imad
AU - Tholakanahalli, Venkatakrishna N
AU - Roukoz, Henri
AU - Knoper, Ryan C
AU - Verma, Varun
AU - Isac, Divya
AU - Manicka, Yatheendhar
N1 - Publisher Copyright:
© The Author(s), under exclusive licence to Springer Science+Business Media, LLC, part of Springer Nature 2024.
PY - 2024
Y1 - 2024
N2 - Background: Traditional cardiac pacemakers commonly have a range of complications related to the presence of intracardiac leads. A new class of extravascular and leadless pacemakers has recently emerged with the potential to mitigate these complications and expand access to cardiac pacing. The objective of this study is to evaluate the implantation, short-term chronic safety, and performance of a novel subxiphoidal extracardiac pacemaker. Methods: Normal Yorkshire Cross swine (n = 16) were implanted with the subxiphoidal pacemaker. The pacemaker was inserted through a midline chest incision and clipped to the underside of the sternum, with the stimulation electrode placed on the anterior pericardium. Animals were chronically paced and followed for 90 days post-implant, with periodic measurement of pacing capture threshold (PCT) and electrode impedance. Results: All 16 animals were successfully implanted with the study device. At implant, a consistent average PCT of 2.2 ± 0.4 V at a pulse width of 1.0 ms was observed in all animals, with an average implant impedance of 648 ± 44 Ω. Chronic pacing was programmed at a rate of 60 bpm, an amplitude of 3.4 ± 0.7 V, and a pulse width of 1.0 ms. PCT rose to 4.6 ± 0.8 V at 14 days and stabilized; at 90 days, PCT was 3.8 ± 1.2 V and electrode impedance was 533 ± 105 Ω. All implanted animals completed the study with no clinically significant findings, no clinically significant abnormalities, and with no adverse events that affected animal welfare. Conclusions: This study demonstrated the safety and feasibility of a novel subxiphoidal extracardiac pacemaker to deliver short-term chronic extravascular therapy. Further studies are required to assess the safety, feasibility, and long-term chronic pacing performance in human subjects.
AB - Background: Traditional cardiac pacemakers commonly have a range of complications related to the presence of intracardiac leads. A new class of extravascular and leadless pacemakers has recently emerged with the potential to mitigate these complications and expand access to cardiac pacing. The objective of this study is to evaluate the implantation, short-term chronic safety, and performance of a novel subxiphoidal extracardiac pacemaker. Methods: Normal Yorkshire Cross swine (n = 16) were implanted with the subxiphoidal pacemaker. The pacemaker was inserted through a midline chest incision and clipped to the underside of the sternum, with the stimulation electrode placed on the anterior pericardium. Animals were chronically paced and followed for 90 days post-implant, with periodic measurement of pacing capture threshold (PCT) and electrode impedance. Results: All 16 animals were successfully implanted with the study device. At implant, a consistent average PCT of 2.2 ± 0.4 V at a pulse width of 1.0 ms was observed in all animals, with an average implant impedance of 648 ± 44 Ω. Chronic pacing was programmed at a rate of 60 bpm, an amplitude of 3.4 ± 0.7 V, and a pulse width of 1.0 ms. PCT rose to 4.6 ± 0.8 V at 14 days and stabilized; at 90 days, PCT was 3.8 ± 1.2 V and electrode impedance was 533 ± 105 Ω. All implanted animals completed the study with no clinically significant findings, no clinically significant abnormalities, and with no adverse events that affected animal welfare. Conclusions: This study demonstrated the safety and feasibility of a novel subxiphoidal extracardiac pacemaker to deliver short-term chronic extravascular therapy. Further studies are required to assess the safety, feasibility, and long-term chronic pacing performance in human subjects.
KW - Arrhythmias
KW - Bradycardia
KW - Cardiac pacing
KW - Leadless pacing
KW - Pacemaker
UR - http://www.scopus.com/inward/record.url?scp=85186476157&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85186476157&partnerID=8YFLogxK
U2 - 10.1007/s10840-024-01778-y
DO - 10.1007/s10840-024-01778-y
M3 - Article
C2 - 38427181
AN - SCOPUS:85186476157
SN - 1383-875X
JO - Journal of Interventional Cardiac Electrophysiology
JF - Journal of Interventional Cardiac Electrophysiology
ER -