TY - JOUR
T1 - Safety and efficacy of the MitraClip® system for severe mitral regurgitation
T2 - A Systematic Review
AU - Vakil, Kairav
AU - Roukoz, Henri
AU - Sarraf, Mohammad
AU - Krishnan, Balaji
AU - Reisman, Mark
AU - Levy, Wayne C.
AU - Adabag, Selcuk
PY - 2014/7/1
Y1 - 2014/7/1
N2 - Background The MitraClip® system is a newer percutaneous device that has shown promising results but data on its safety and efficacy in low- and high-surgical risk populations continues to evolve. We performed a systematic review of the published studies reporting the safety and efficacy of MitraClip® implantation for treatment of moderate to severe and severe mitral regurgitation (MR). Methods Reviewers independently searched for relevant articles in Medline and abstracted clinical information based on pre-defined criteria and end-points. Patients were classified as low- or high-surgical risk for conventional mitral valve (MV) surgery based on Society of Thoracic Surgeons score, EuroSCORE, or surgeon discretion. Primary safety outcome was 30-day mortality and primary efficacy outcomes were freedom from death, ≥3+ MR, and requirement for MV surgery during follow-up. Results Immediate and long-term outcomes of 16 studies, including 2980 patients (age 73.7 ± 0.6 years; 63.3% males) receiving the MitraClip®, were analyzed. Acute procedural success rate was 91.4% with a procedural mortality of 0.1%. Mortality at 30 days and long-term (310 days) follow-up was 4.2 and 15.8%, respectively and was significantly higher in the high-risk group (P = 0.003 and 0.019, respectively). Incidence of major procedural adverse outcomes was relatively low with blood transfusion accounting for most events. At follow-up, the number of patients with ≥3+ MR reduced from 96.3% to 14.7% (P < 0.001), and those with NYHA class III/IV reduced from 83.2% to 23.4% (P < 0.001). Conclusion MitraClip® implantation for moderate to severe or severe MR appears to be safe with a very low procedural mortality. There is significant improvement in functional outcomes although long-term mortality is high, especially in high surgical risk patients.
AB - Background The MitraClip® system is a newer percutaneous device that has shown promising results but data on its safety and efficacy in low- and high-surgical risk populations continues to evolve. We performed a systematic review of the published studies reporting the safety and efficacy of MitraClip® implantation for treatment of moderate to severe and severe mitral regurgitation (MR). Methods Reviewers independently searched for relevant articles in Medline and abstracted clinical information based on pre-defined criteria and end-points. Patients were classified as low- or high-surgical risk for conventional mitral valve (MV) surgery based on Society of Thoracic Surgeons score, EuroSCORE, or surgeon discretion. Primary safety outcome was 30-day mortality and primary efficacy outcomes were freedom from death, ≥3+ MR, and requirement for MV surgery during follow-up. Results Immediate and long-term outcomes of 16 studies, including 2980 patients (age 73.7 ± 0.6 years; 63.3% males) receiving the MitraClip®, were analyzed. Acute procedural success rate was 91.4% with a procedural mortality of 0.1%. Mortality at 30 days and long-term (310 days) follow-up was 4.2 and 15.8%, respectively and was significantly higher in the high-risk group (P = 0.003 and 0.019, respectively). Incidence of major procedural adverse outcomes was relatively low with blood transfusion accounting for most events. At follow-up, the number of patients with ≥3+ MR reduced from 96.3% to 14.7% (P < 0.001), and those with NYHA class III/IV reduced from 83.2% to 23.4% (P < 0.001). Conclusion MitraClip® implantation for moderate to severe or severe MR appears to be safe with a very low procedural mortality. There is significant improvement in functional outcomes although long-term mortality is high, especially in high surgical risk patients.
KW - MitraClip
KW - edge-to-edge mitral valve repair
KW - mitral clip, mitral regurgitation
UR - http://www.scopus.com/inward/record.url?scp=84903515434&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=84903515434&partnerID=8YFLogxK
U2 - 10.1002/ccd.25347
DO - 10.1002/ccd.25347
M3 - Review article
C2 - 24323764
AN - SCOPUS:84903515434
SN - 1522-1946
VL - 84
SP - 129
EP - 136
JO - Catheterization and Cardiovascular Interventions
JF - Catheterization and Cardiovascular Interventions
IS - 1