Safety and effectiveness of mentor's memorygel implants at 6 years

Bruce L Cunningham, Jonathan McCue

Research output: Contribution to journalArticlepeer-review

159 Scopus citations

Abstract

Background: In November 2006, the FDA approved the Premarket Approval PMA applications for the round, cohesive, silicone gel-filled breast implants of Mentor (MemoryGel) and Allergan. Since that time, the devices have been widely available to plastic surgeons and their use is rapidly eclipsing that of the saline breast implants. Patients in the Core clinical studies supporting these approvals continue to be followed through for 10 years, with comprehensive annual patient and physician-reported evaluations of safety and efficacy. Methods: One thousand and eight (1,008) female patients had data collected on 1,898 implants, and were enrolled at 48 sites. Key complication rates were recorded with Kaplan-Meier estimated cumulative incidence calculation for each. Results: Rupture rate, suspected and confirmed, for primary augmentation was 1.1% (95% CI, 0.3-4.3), and that for primary reconstruction patients was 3.8% (95% CI, 1.4-9.8). Capsular contracture rates for clinically significant Baker III/IV contracture for primary augmentation was 9.8% (95% C I, 7.6-12.7), and that for primary reconstruction was 13.7% (95% CI, 9.7-19.1). The reoperation incidence for primary augmentation and primary reconstruction was 19.4 and 33.9%, respectively, with explantation and replacement with a study device in 3.9% of primary augmentations and 10.4% of primary reconstructions. Conclusions: Mentor MemoryGel Silicone Breast implants represent a safe and effective choice for women seeking breast augmentation or breast reconstruction following mastectomy.

Original languageEnglish (US)
Pages (from-to)440-444
Number of pages5
JournalAesthetic Plastic Surgery
Volume33
Issue number3
DOIs
StatePublished - May 2009

Keywords

  • Complications
  • Effectiveness
  • Follow-up
  • Mentor
  • Outcomes
  • Rupture
  • Safety
  • Silicone breast implants

Fingerprint

Dive into the research topics of 'Safety and effectiveness of mentor's memorygel implants at 6 years'. Together they form a unique fingerprint.

Cite this