The effect of loading rate on specimen calibration was investigated for an implantable force sensor of the two-point loading variety. This variety of sensor incorporates a strain gage to measure the compressive load applied to the sensor due to tensile loading in a soft tissue specimen. The Achilles tendon in each of four human cadaveric lower extremities was instrumented with a force sensor and then loaded in tension using a materials testing machine. Each specimen was tensile tested at three different displacement rates, 0.25, 2.5 and 12.7 cm s-1, corresponding with mean loading rates of 33.8, 513.2, and 2838.6 N s-1, respectively. A calibration curve relating the force sensor signal and applied tendon tension was generated for each specimen/displacement rate combination. For each specimen, calibration curves were compared by calculating an RMS error for the entire data set (ε(RMS) = 1.6% of the full toad value) and a coefficient of determination, R2, of a curve fit through all of the data (R2 = 99.6%). Over the range of rates tested, no measurable change in sensor sensitivity due to loading rate was observed. Hysteresis for all displacement rates was on the order of 2.4%.
| Original language | English (US) |
|---|---|
| Pages (from-to) | 203-207 |
| Number of pages | 5 |
| Journal | Journal of Biomechanics |
| Volume | 32 |
| Issue number | 2 |
| DOIs | |
| State | Published - Feb 1 1999 |
Keywords
- Calibration
- Force
- Implant
- Tendon
- Transducer
Fingerprint
Dive into the research topics of 'Rate-independent characteristics of an arthroscopically implantable force probe in the human achilles tendon'. Together they form a unique fingerprint.Cite this
- APA
- Standard
- Harvard
- Vancouver
- Author
- BIBTEX
- RIS