Abstract
A prototype 64-element, 75° cylindrical-section ultrasonic phased-array hyperthermia applicator has been designed and constructed. The ability of this applicator to focus ultrasonic energy at its geometric focus was verified in a water medium. The array was then driven by excitation vectors obtained using the pseudoinverse pattern synthesis method to generate shifted-focus and double-focus field patterns. The measured intensity profiles agree well with the theoretically predicted ones.
Original language | English (US) |
---|---|
Pages (from-to) | 995-998 |
Number of pages | 4 |
Journal | Ultrasonics Symposium Proceedings |
Volume | 2 |
State | Published - 1989 |
Event | IEEE 1989 Ultrasonics Symposium - Montreal, Que, Can Duration: Oct 3 1989 → Oct 6 1989 |