N2 - The aim of this study is to investigate the safety and effectiveness of Excimer laser-assisted lead extraction. Consecutive lead extraction in 16 patients using the excimer laser sheath from Veterans Administration Medical Center, University of Minnesota between March 2008 and December 2009 were analyzed retrospectively. The clinical characteristics including basic disease, cardiac function, indication and reason for lead removal, lead characteristic, and complication were summarized. Thirty one leads (laser sheaths were used in 25 leads) in 16 patients [mean age (67.4 ± 9.1) years, range 54 - 82 years] were extracted. All leads were extracted successfully without complication. Most patients had severe heart diseases with left ventricular ejection fraction of 20% - 60% (mean 36.4% ± 13.4%). Mean implant time was (73.9 ± 62.9) months (range 3 - 251 months). Nine patients had Class I indications for lead extraction, 7 patients for Class II indications. Atrial lead fractured in one patient during lead extraction, the remaining part of the lead was taken out from femoral vein by a snare-catheter apparatus. Another patient developed CS dissection and small pericardium perfusion during LV lead implantation (upgrade to BiV-ICD) post successful lead extraction and patient discharged without complication. Our single center experience suggests that Excimer Laser Sheath Lead Removal System is safe and effective in extracting pacemaker and ICD leads.
AB - The aim of this study is to investigate the safety and effectiveness of Excimer laser-assisted lead extraction. Consecutive lead extraction in 16 patients using the excimer laser sheath from Veterans Administration Medical Center, University of Minnesota between March 2008 and December 2009 were analyzed retrospectively. The clinical characteristics including basic disease, cardiac function, indication and reason for lead removal, lead characteristic, and complication were summarized. Thirty one leads (laser sheaths were used in 25 leads) in 16 patients [mean age (67.4 ± 9.1) years, range 54 - 82 years] were extracted. All leads were extracted successfully without complication. Most patients had severe heart diseases with left ventricular ejection fraction of 20% - 60% (mean 36.4% ± 13.4%). Mean implant time was (73.9 ± 62.9) months (range 3 - 251 months). Nine patients had Class I indications for lead extraction, 7 patients for Class II indications. Atrial lead fractured in one patient during lead extraction, the remaining part of the lead was taken out from femoral vein by a snare-catheter apparatus. Another patient developed CS dissection and small pericardium perfusion during LV lead implantation (upgrade to BiV-ICD) post successful lead extraction and patient discharged without complication. Our single center experience suggests that Excimer Laser Sheath Lead Removal System is safe and effective in extracting pacemaker and ICD leads.
UR - http://www.scopus.com/inward/record.url?scp=84855184380&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=84855184380&partnerID=8YFLogxK
M3 - Article
C2 - 21624315
AN - SCOPUS:84855184380
SN - 0253-3758
VL - 39
SP - 363
EP - 368
JO - Zhonghua xin xue guan bing za zhi
JF - Zhonghua xin xue guan bing za zhi
IS - 4
ER -