Non-invasive studies of peripheral vascular compliance using a non-occluding photoplethysmographic method

E. A. López-Beltrán, P. L. Blackshear, Stanley M Finkelstein, Jay N Cohn

Research output: Contribution to journalArticle

47 Scopus citations

Abstract

A non-invasive technique is implemented to measure a peripheral vascular compliance index Cindex, using an infrared photoplethysmographic waveform as an indicator of intravascular volume change and a continuous blood pressure monitor to measure the blood pressure during each heart-beat. The non-linear behaviour of Cindex with pressure and the effect of age on Cindex are studied in 62 males (15-73 years). Repeatability tests and the effect of ice-water exposure of a portion of a limb are studied in 10 and 14 subjects, respectively. For each individual, Cindex measurements are taken at discrete values of local mean arterial pressure (Pmean), and a Cindex against Pmean plot is obtained. There is a statistically significant difference (p<0.05) in Cindex tor the lower values of Pmean (60-100 mmHg) between two age groups formed (15-52 and 58-73 years). The cold-pressor test (CPT) shows a 68% median decrease in Cindex, with an inter-quartile range of 60-77%, in a matter of seconds. The results suggest that Cindex may be a useful noninvasive indicator of peripheral vascular compliance in humans.

Original languageEnglish (US)
Pages (from-to)748-753
Number of pages6
JournalMedical and Biological Engineering and Computing
Volume36
Issue number6
DOIs
StatePublished - Jan 1 1998

Keywords

  • Age-effect
  • Non-invasive compliance
  • Peripheral compliance
  • Photoplethysmography
  • Pressure-dependent compliance
  • Repeatability

Fingerprint Dive into the research topics of 'Non-invasive studies of peripheral vascular compliance using a non-occluding photoplethysmographic method'. Together they form a unique fingerprint.

  • Cite this