TY - JOUR
T1 - Microgap Formation in Conical Implant-Abutment Connections Under Oblique Loading
T2 - Influence of Cone Angle Mismatch Through Finite Element Analysis
AU - Gantier-Takano, Marlene Kasumi
AU - Xing, Yiyun
AU - Ye, Ning
AU - Aparicio, Conrado
AU - Navarro Cuéllar, Carlos
AU - Meira, Josete Barbosa Cruz
AU - Fok, Alex Siu Lun
N1 - Publisher Copyright:
© 2025 Wiley Periodicals LLC.
PY - 2025/2
Y1 - 2025/2
N2 - Objectives: This study evaluated different designs of the conical implant-abutment connection (IAC) and their resistance to microgap formation under oblique loads as specified by the ISO standard for testing dental implants. Also evaluated was the effect of deviations from the ISO specifications on the outcomes. Methods: Finite element analysis was conducted to compare the microgap formation and stress distribution among three conical IAC designs (A, B, and C) in two loading configurations: one compliant with ISO 14801 and one with a modified load adaptor (non-ISO). The different IAC designs varied in the taper, diameter, and cone height. The cone angle mismatch (Cam) between the implant and abutment was considered. A torque of 20 Ncm and oblique loads (up to 400 N) were simulated. Results: The stresses produced by the screw-tightening torque varied among the different IAC designs. The contact height was approximately 0.3 mm for Designs A and B, and less than 0.03 mm for Design C. Under oblique loads, Design A maintained IAC sealing without gap formation up to 400 N. With the ISO adaptor, gaps appeared in Design B at 300 N and in Design C at 90 N. The non-ISO adaptor resulted in gap formation at 160 N for Design B and at 50 N for Design C. Conclusions: The IAC design and cone angle mismatch significantly influenced microgap formation, with some designs showing zero gaps even when the oblique load reached 400 N. The non-ISO adaptor increased gap formation in IACs B and C.
AB - Objectives: This study evaluated different designs of the conical implant-abutment connection (IAC) and their resistance to microgap formation under oblique loads as specified by the ISO standard for testing dental implants. Also evaluated was the effect of deviations from the ISO specifications on the outcomes. Methods: Finite element analysis was conducted to compare the microgap formation and stress distribution among three conical IAC designs (A, B, and C) in two loading configurations: one compliant with ISO 14801 and one with a modified load adaptor (non-ISO). The different IAC designs varied in the taper, diameter, and cone height. The cone angle mismatch (Cam) between the implant and abutment was considered. A torque of 20 Ncm and oblique loads (up to 400 N) were simulated. Results: The stresses produced by the screw-tightening torque varied among the different IAC designs. The contact height was approximately 0.3 mm for Designs A and B, and less than 0.03 mm for Design C. Under oblique loads, Design A maintained IAC sealing without gap formation up to 400 N. With the ISO adaptor, gaps appeared in Design B at 300 N and in Design C at 90 N. The non-ISO adaptor resulted in gap formation at 160 N for Design B and at 50 N for Design C. Conclusions: The IAC design and cone angle mismatch significantly influenced microgap formation, with some designs showing zero gaps even when the oblique load reached 400 N. The non-ISO adaptor increased gap formation in IACs B and C.
KW - dental implant-abutments design
KW - dimensional measurement accuracy
KW - finite element analysis
KW - mechanical
KW - peri-implantitis
KW - stress
UR - http://www.scopus.com/inward/record.url?scp=85215787893&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85215787893&partnerID=8YFLogxK
U2 - 10.1111/cid.13436
DO - 10.1111/cid.13436
M3 - Article
C2 - 39840917
AN - SCOPUS:85215787893
SN - 1523-0899
VL - 27
JO - Clinical Implant Dentistry and Related Research
JF - Clinical Implant Dentistry and Related Research
IS - 1
M1 - e13436
ER -