Abstract
We describe the design, fabrication, and use of a radio frequency reflectometer integrated with a microfluidic system, applied to the very high-throughput measurement of micron-scale particles, passing in a microfluidic channel through the sensor region. The device operates as a microfabricated Coulter counter [U.S. Patent No. 2656508 (1953)], similar to a design we have described previously, but here with significantly improved electrode geometry as well as including electronic tuning of the reflectometer; the two improvements yielding an improvement by more than a factor of 10 in the signal to noise and in the diametric discrimination of single particles. We demonstrate the high-throughput discrimination of polystyrene beads with diameters in the 4-10 μm range, achieving diametric resolutions comparable to the intrinsic spread of diameters in the bead distribution, at rates in excess of 15 × 10 6 beads/h.
| Original language | English (US) |
|---|---|
| Article number | 104301 |
| Journal | Review of Scientific Instruments |
| Volume | 78 |
| Issue number | 10 |
| DOIs | |
| State | Published - 2007 |
Bibliographical note
Funding Information:This work was supported in part by the NIH through an NIH PEN grant and by the DMEA/DARPA Center for Nanoscience Innovation for Defence.
Fingerprint
Dive into the research topics of 'Microfabricated high-throughput electronic particle detector'. Together they form a unique fingerprint.Cite this
- APA
- Standard
- Harvard
- Vancouver
- Author
- BIBTEX
- RIS