Skip to main navigation Skip to search Skip to main content

Microfabricated high-throughput electronic particle detector

Research output: Contribution to journalArticlepeer-review

Abstract

We describe the design, fabrication, and use of a radio frequency reflectometer integrated with a microfluidic system, applied to the very high-throughput measurement of micron-scale particles, passing in a microfluidic channel through the sensor region. The device operates as a microfabricated Coulter counter [U.S. Patent No. 2656508 (1953)], similar to a design we have described previously, but here with significantly improved electrode geometry as well as including electronic tuning of the reflectometer; the two improvements yielding an improvement by more than a factor of 10 in the signal to noise and in the diametric discrimination of single particles. We demonstrate the high-throughput discrimination of polystyrene beads with diameters in the 4-10 μm range, achieving diametric resolutions comparable to the intrinsic spread of diameters in the bead distribution, at rates in excess of 15 × 10 6 beads/h.

Original languageEnglish (US)
Article number104301
JournalReview of Scientific Instruments
Volume78
Issue number10
DOIs
StatePublished - 2007

Bibliographical note

Funding Information:
This work was supported in part by the NIH through an NIH PEN grant and by the DMEA/DARPA Center for Nanoscience Innovation for Defence.

Fingerprint

Dive into the research topics of 'Microfabricated high-throughput electronic particle detector'. Together they form a unique fingerprint.

Cite this