Abstract
We present a case of a 13-year old patient with partially reversibly post-operative heart block who underwent leadless pacemaker placement. After post-anesthesia wretching/gagging episode she developed device microdislodgement and increased/intolerable capture thresholds. The device was removed and another placed with adequate thresholds for good longevity.
| Original language | English (US) |
|---|---|
| Pages (from-to) | 132-134 |
| Number of pages | 3 |
| Journal | Indian Pacing and Electrophysiology Journal |
| Volume | 20 |
| Issue number | 3 |
| DOIs | |
| State | Published - May 1 2020 |
Bibliographical note
Publisher Copyright:© 2020 Indian Heart Rhythm Society
Keywords
- Leadless pacemaker removal
- Pediatric