Mechanical and spectroscopic analysis of retrieved/failed dental implants

  • Umer Daood
  • , Ninette Banday
  • , Zohaib Akram
  • , James K.H. Tsoi
  • , Prasanna Neelakantan
  • , Hanan Omar
  • , Tariq Abduljabbar
  • , Fahim Vohra
  • , Nawwaf Al-Hamoudi
  • , Amr S. Fawzy

Research output: Contribution to journalArticlepeer-review

Abstract

The purpose of this study was to examine surface alterations and bone formation on the surface of failed dental implants (Straumann [ST] and TiUnite [TiUn]) removed due to any biological reason. In addition, failure analysis was performed to test mechanical properties. Dental implants (n = 38) from two manufacturers were collected and subjected to chemical cleaning. The presence of newly formed hydroxyapatite bone around failed implants was evaluated using micro-Raman spectroscopy. Scanning electron microscopy was used to identify surface defects. Mechanical testing was performed using a Minneapolis servo-hydraulic system (MTS) along with indentation using a universal testing machine and average values were recorded. A statistical analysis of mechanical properties was done using an unpaired t test, and correlation between observed defects was evaluated using Chi-square (p = 0.05). Apatite-formation was evident in both implants, but was found qualitatively more in the ST group. No significant difference was found in indentation between the two groups (p > 0.05). The percentage of "no defects" was significantly lower in the ST group (71%). Crack-like and full-crack defects were observed in 49% and 39% of TiUn. The ST group showed 11,061 cycles to failure as compared with 10,021 cycles in the TiUnite group. Implant failure mechanisms are complex with a combination of mechanical and biological reasons and these factors are variable with different implant systems.

Original languageEnglish (US)
Article number201
JournalCoatings
Volume7
Issue number11
DOIs
StatePublished - Nov 1 2017
Externally publishedYes

Bibliographical note

Publisher Copyright:
© 2017 by the authors.

Keywords

  • Dental implant
  • Failed
  • Hydroxyapatite
  • Mechanical
  • Nanoindentation
  • Raman
  • Titanium

Fingerprint

Dive into the research topics of 'Mechanical and spectroscopic analysis of retrieved/failed dental implants'. Together they form a unique fingerprint.

Cite this