Abstract
Purpose: Suprachoroidal hemorrhage (SCH) is a rare but vision-threatening complication in patients with keratoprosthesis devices (KPro), particularly in the setting of concurrent glaucoma tube shunts. Although there are many approaches to draining an SCH, surgery can be especially challenging in these patients because a crowded anterior chamber, and frequent anterior extension of the SCH. Methods: A case report is discussed. Results: We describe a novel approach to surgical drainage of SCH in a 64-year-old monocular patient with a Type I Boston KPro, an Ahmed valve, and aphakia. Conclusion: Successful repair of appositional SCH in the context of KPro can be safely and effectively achieved by injecting viscoelastic through the backplate holes of the KPro, thereby providing internal tamponade within the vitreous compartment while liquefied hemorrhage is drained by sclerotomy.
Original language | English (US) |
---|---|
Pages (from-to) | 210-213 |
Number of pages | 4 |
Journal | Journal of VitreoRetinal Diseases |
Volume | 6 |
Issue number | 3 |
DOIs | |
State | Published - May 1 2022 |
Bibliographical note
Publisher Copyright:© The Author(s) 2021.
Keywords
- keratoprosthesis
- suprachoroidal hemorrhage
- viscoelastic
PubMed: MeSH publication types
- Journal Article