T1 - Is the Implant in Bone? The Accuracy of CT and Fluoroscopic Imaging for Detecting Malpositioned Pelvic Screw and SI Fusion Implants
AU - San Miguel-Ruiz, Jose E.
AU - Polly, David
AU - Albersheim, Melissa
AU - Sembrano, Jonathan
AU - Takahashi, Takashi
AU - Lender, Paul
AU - Martin, Christopher T.
N1 - Publisher Copyright:
Copyright © The Iowa Orthopaedic Journal 2021.
PY - 2021/1/1
Y1 - 2021/1/1
N2 - Background: Spine fusions to the pelvis have been associated with increased strain to the sacroiliac joint (SI) and possibly continued postoperative low back pain. To minimize this, concomitant SI joint fusion at the time of lumbopelvic fixation has been advocated. This requires concomitant placement of sacral alar iliac screws (S2AI) for lumbopelvic fixation and triangular titanium rods (TTR) for the SI joint fusion. Traditionally, surgeons have mostly relied on fluoroscopic images to confirm final implant position and patient safety after pelvic instrumentation, although computer tomography (CT) has also been used. Methods: We wanted to know which imaging modality, if any, was superior in helping to identify malpositioned implants during concomitant lumbopelvic fixation and SI joint fusion. We instrumented pelvic sawbones models with S2AI screws, TTR's, or both in the correct anatomic positions or malpositioned variants that led to known cortical breaches. Pelvic models were then imaged with fluoroscopy and CT, and the images assessed by blinded reviewers (spine surgeons and a musculoskeletal radiologist) for the presence of cortical breaches, the identity of the breached implant, and its direction. The responses of the blinded reviewers were then compared to the known position of the implants and Kappa coefficient calculated to determine agreement. Results: We found that thorough evaluation of implant position with multiple fluoroscopic views (kappa 0.641) or CT imaging (kappa 0.906) allowed reviewers to assess implant position, identity, and breach direction. Conclusion: These findings suggest that intraoperative CT imaging allows surgeons to make the best decision regarding implant position prior to leaving the operating room, thus potentially improving patient safety and unplanned returns to the operating room.Level of Evidence: V.
AB - Background: Spine fusions to the pelvis have been associated with increased strain to the sacroiliac joint (SI) and possibly continued postoperative low back pain. To minimize this, concomitant SI joint fusion at the time of lumbopelvic fixation has been advocated. This requires concomitant placement of sacral alar iliac screws (S2AI) for lumbopelvic fixation and triangular titanium rods (TTR) for the SI joint fusion. Traditionally, surgeons have mostly relied on fluoroscopic images to confirm final implant position and patient safety after pelvic instrumentation, although computer tomography (CT) has also been used. Methods: We wanted to know which imaging modality, if any, was superior in helping to identify malpositioned implants during concomitant lumbopelvic fixation and SI joint fusion. We instrumented pelvic sawbones models with S2AI screws, TTR's, or both in the correct anatomic positions or malpositioned variants that led to known cortical breaches. Pelvic models were then imaged with fluoroscopy and CT, and the images assessed by blinded reviewers (spine surgeons and a musculoskeletal radiologist) for the presence of cortical breaches, the identity of the breached implant, and its direction. The responses of the blinded reviewers were then compared to the known position of the implants and Kappa coefficient calculated to determine agreement. Results: We found that thorough evaluation of implant position with multiple fluoroscopic views (kappa 0.641) or CT imaging (kappa 0.906) allowed reviewers to assess implant position, identity, and breach direction. Conclusion: These findings suggest that intraoperative CT imaging allows surgeons to make the best decision regarding implant position prior to leaving the operating room, thus potentially improving patient safety and unplanned returns to the operating room.Level of Evidence: V.
KW - SI joint fusion
KW - intraoperative CT
KW - sacral alar 2 iliac screws
KW - spinopelvic fixation
KW - triangular titanium rods
UR - http://www.scopus.com/inward/record.url?scp=85116830422&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85116830422&partnerID=8YFLogxK
M3 - Article
C2 - 34552409
AN - SCOPUS:85116830422
SN - 1541-5457
VL - 41
SP - 89
EP - 94
JO - The Iowa orthopaedic journal
JF - The Iowa orthopaedic journal
IS - 1
ER -