TY - JOUR
T1 - Ionic liquid-based reference electrodes for miniaturized ion sensors
T2 - What can go wrong?
AU - Mousavi, Maral P.S.
AU - Abd El-Rahman, Mohamed K.
AU - Tan, Edward K.W.
AU - Sigurslid, Haakon H.
AU - Arkan, Nooralhuda
AU - Lane, John S.
AU - Whitesides, George M.
AU - Bühlmann, Philippe
N1 - Publisher Copyright:
© 2019 Elsevier B.V.
PY - 2019/12/12
Y1 - 2019/12/12
N2 - Ionic liquid-based reference electrodes, especially those containing 1-octyl-3-methylimidazolium bis(trifluoromethylsulfonyl)imide (MeOctIm TFSI), are increasingly used in potentiometric measurements. They replace the conventional salt bridges and establish the reference potential through local partitioning of the ions of the ionic liquid between the reference membrane and the sample solution. Unless the electrochemical cell and the measurement protocol are designed appropriately, these ions can interfere with the response of ion-selective electrodes (ISEs) with polymeric membranes. This work characterizes the effect of MeOctIm TFSI on the response of K+, Na+, and Cl− ISEs with polymeric membranes. The leaching of MeOctIm TFSI from the reference membrane to the sample solution was monitored with UV–vis spectroscopy. The concentration of MeOctIm TFSI in the aqueous phase increased gradually and plateaued at approximately 1.0 mM. Concentrations as low as 100 μM of MeOctIm TFSI caused large changes to the emf (50–150 mV) of K+, Na+, and Cl− ISEs. The presence of 10 μM of TFSI− was enough to cause Donnan failure of K+ ISEs with valinomycin as ionophore (that is, co-extraction of K+ and TFSI− into the sensing membrane). Use of less lipophilic anions such as tetrafluoroborate (BF4-) or triflate (OTf-) postponed the onset of Donnan failure of cation-selective ISEs to higher concentrations of the anion, but decreased the stability in the reference potential and lifetime of the reference electrode. These results imply that although MeOctIm TFSI-based reference electrodes provide sample-independent and stable electrical potentials, they should be used with caution for measurements with polymeric-membrane ISEs, due to strong interference of both MeOctIm+ and TFSI− with measured values of emf.
AB - Ionic liquid-based reference electrodes, especially those containing 1-octyl-3-methylimidazolium bis(trifluoromethylsulfonyl)imide (MeOctIm TFSI), are increasingly used in potentiometric measurements. They replace the conventional salt bridges and establish the reference potential through local partitioning of the ions of the ionic liquid between the reference membrane and the sample solution. Unless the electrochemical cell and the measurement protocol are designed appropriately, these ions can interfere with the response of ion-selective electrodes (ISEs) with polymeric membranes. This work characterizes the effect of MeOctIm TFSI on the response of K+, Na+, and Cl− ISEs with polymeric membranes. The leaching of MeOctIm TFSI from the reference membrane to the sample solution was monitored with UV–vis spectroscopy. The concentration of MeOctIm TFSI in the aqueous phase increased gradually and plateaued at approximately 1.0 mM. Concentrations as low as 100 μM of MeOctIm TFSI caused large changes to the emf (50–150 mV) of K+, Na+, and Cl− ISEs. The presence of 10 μM of TFSI− was enough to cause Donnan failure of K+ ISEs with valinomycin as ionophore (that is, co-extraction of K+ and TFSI− into the sensing membrane). Use of less lipophilic anions such as tetrafluoroborate (BF4-) or triflate (OTf-) postponed the onset of Donnan failure of cation-selective ISEs to higher concentrations of the anion, but decreased the stability in the reference potential and lifetime of the reference electrode. These results imply that although MeOctIm TFSI-based reference electrodes provide sample-independent and stable electrical potentials, they should be used with caution for measurements with polymeric-membrane ISEs, due to strong interference of both MeOctIm+ and TFSI− with measured values of emf.
KW - Ion sensing
KW - Ion-selective electrode
KW - Ionic liquid Salt Bridge
KW - Liquid junction potential
KW - Potentiometry
KW - Reference electrode
UR - http://www.scopus.com/inward/record.url?scp=85072250719&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85072250719&partnerID=8YFLogxK
U2 - 10.1016/j.snb.2019.127112
DO - 10.1016/j.snb.2019.127112
M3 - Article
AN - SCOPUS:85072250719
SN - 0925-4005
VL - 301
JO - Sensors and Actuators, B: Chemical
JF - Sensors and Actuators, B: Chemical
M1 - 127112
ER -