TY - JOUR
T1 - Ion-Selective Electrodes
T2 - Selectivity Coefficients for Interfering Ions of the Opposite Charge Sign
AU - Honig, Madeline L.
AU - Bühlmann, Philippe
N1 - Publisher Copyright:
© 2025 American Chemical Society.
PY - 2025
Y1 - 2025
N2 - The way upper limits of detection (LODs) are typically reported in the ion-selective electrode (ISE) literature is unfortunately outdated. It is well understood that the upper LOD of a polymeric-membrane ISE is limited by Donnan failure, that is, the transfer of primary ions along with interfering ions of the opposite charge sign (commonly referred to as counterions) from the sample into the sensing membrane. However, it is often difficult to compare upper LODs for ISEs from different sources. The majority of publications on ISEs describe Donnan failure for one type of counterion only, making it impossible for end users to predict the interference for other counterions. Moreover, linear ranges for ISEs based on different ionophores cannot be compared to one another when Donnan failure was reported for different counterions. To this end, we introduce here selectivity coefficients, KI,XpotX, for interfering counterions. Using this new concept, the primary ion activity at which Donnan failure occurs can be readily predicted from measured KI,XpotX values by the use of the uncomplicated expression aXzI/zx/KI,XpotX. Consistent with the intuition that many ISE users have for conventional selectivity coefficients, large KI,XpotX values are characteristic for counterions that interfere strongly. We show experimentally that trends as predicted by the phase boundary model for Donnan failure, such as the effects of counterion hydrophobicity and ionophore complex stability, are often accurately predicted with the KI,XpotX approach. However, there are notable exceptions when the underlying assumptions made by users do not apply, such as when counterions unexpectedly form aggregates with other species in the sensing membranes. The empirically measured KI,XpotX coefficients enable the discovery of such phenomena, opening a rational path to improving upper LODs and, thereby, linear response ranges.
AB - The way upper limits of detection (LODs) are typically reported in the ion-selective electrode (ISE) literature is unfortunately outdated. It is well understood that the upper LOD of a polymeric-membrane ISE is limited by Donnan failure, that is, the transfer of primary ions along with interfering ions of the opposite charge sign (commonly referred to as counterions) from the sample into the sensing membrane. However, it is often difficult to compare upper LODs for ISEs from different sources. The majority of publications on ISEs describe Donnan failure for one type of counterion only, making it impossible for end users to predict the interference for other counterions. Moreover, linear ranges for ISEs based on different ionophores cannot be compared to one another when Donnan failure was reported for different counterions. To this end, we introduce here selectivity coefficients, KI,XpotX, for interfering counterions. Using this new concept, the primary ion activity at which Donnan failure occurs can be readily predicted from measured KI,XpotX values by the use of the uncomplicated expression aXzI/zx/KI,XpotX. Consistent with the intuition that many ISE users have for conventional selectivity coefficients, large KI,XpotX values are characteristic for counterions that interfere strongly. We show experimentally that trends as predicted by the phase boundary model for Donnan failure, such as the effects of counterion hydrophobicity and ionophore complex stability, are often accurately predicted with the KI,XpotX approach. However, there are notable exceptions when the underlying assumptions made by users do not apply, such as when counterions unexpectedly form aggregates with other species in the sensing membranes. The empirically measured KI,XpotX coefficients enable the discovery of such phenomena, opening a rational path to improving upper LODs and, thereby, linear response ranges.
KW - Donnan failure
KW - counterions
KW - ion-selective electrodes
KW - ionophore
KW - selectivity coefficients
KW - upper detection limit
UR - http://www.scopus.com/inward/record.url?scp=105002014443&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=105002014443&partnerID=8YFLogxK
U2 - 10.1021/acssensors.5c00126
DO - 10.1021/acssensors.5c00126
M3 - Article
C2 - 40186583
AN - SCOPUS:105002014443
SN - 2379-3694
JO - ACS Sensors
JF - ACS Sensors
ER -