Abstract
Utilization of Berlin Excor as a pediatric bridge to transplantation device continues to increase. Thromboembolic complications continue to be an important morbidity related to ventricular assist devices. Obstructions to the device cannulae are rare but potentially life-threatening complications of ventricular assist devices, but inflow cannula obstruction is hitherto unreported. We present a case of a 4-year-old supported with Berlin BiVAD who demonstrated acute decompensation related to thrombus within the left atrial cannula. We describe the diagnosis, intervention, as well as novel rescue using existing cannulae for extracorporeal membrane oxygenation support.
Original language | English (US) |
---|---|
Pages (from-to) | 625-628 |
Number of pages | 4 |
Journal | International Journal of Artificial Organs |
Volume | 43 |
Issue number | 9 |
DOIs | |
State | Published - Sep 1 2020 |
Externally published | Yes |
Bibliographical note
Publisher Copyright:© The Author(s) 2020.
Keywords
- antithrombotic therapy
- apheresis and detoxification techniques
- artificial kidney
- cardiac assist and artificial heart
- Pediatric
- pediatric circulatory support
- pediatric ECMO
- pediatric VAD support
- thromboembolism