Incomplete left atrial appendage occlusion and thrombus formation after Watchman implantation treated with anticoagulation followed by further transcatheter closure with a second-generation Amplatzer Cardiac Plug (Amulet device)

Simon Cheung Chi Lam, Stefan Bertog, Horst Sievert

Research output: Contribution to journalArticlepeer-review

31 Scopus citations

Abstract

We report a case of incomplete left atrial appendage (LAA) closure after Watchman device (Atritech, Boston Scientific, Natrick, MA) implantation which subsequently developed a thrombus 3 years after the initial procedure. The thrombus resolved after a short period of anticoagulation with warfarin, and the LAA was successfully occluded with a second-generation Amplatzer Cardiac Plug (Amulet device, AGA, St Jude Medical, Minneapolis, MN). Incomplete LAA closure may be associated with increased risk of thrombus formation and further closure with a second device after a course of anticoagulation may be a reasonable and feasible strategy.

Original languageEnglish (US)
Pages (from-to)321-327
Number of pages7
JournalCatheterization and Cardiovascular Interventions
Volume85
Issue number2
DOIs
StatePublished - Feb 1 2015
Externally publishedYes

Bibliographical note

Publisher Copyright:
© 2014 Wiley Periodicals, Inc.

Keywords

  • Fistula/shunts
  • Left atrial appendage closure
  • Transseptal cath

Fingerprint

Dive into the research topics of 'Incomplete left atrial appendage occlusion and thrombus formation after Watchman implantation treated with anticoagulation followed by further transcatheter closure with a second-generation Amplatzer Cardiac Plug (Amulet device)'. Together they form a unique fingerprint.