T1 - Impact of very long time output variation in the treatment of total marrow irradiation with helical tomotherapy
AU - Takahashi, Yutaka
AU - Hui, Susanta K.
N1 - Funding Information:
This work was supported by the National Institute of Health grants (1R01CA154491-01).
PY - 2013/5/20
Y1 - 2013/5/20
N2 - Background: Beam-on time in Total Marrow Irradiation (TMI) delivery with helical tomotherapy is more than 30 minutes. The purpose of this study was to investigate extended time output variation in tomotherapy machine without dose servo system and its impact on the dosimetry of TMI planning. Materials and methods: The calibration procedures with 1800 seconds delivery were conducted. The slab and cylindrical phantoms were used for static and rotational output variation measurements, respectively. All measurements were performed in 0.1 second interval with an Exradin A1SL ionization chamber (Standard Imaging Inc., Madison, WI, USA) connected to the tomoelectrometer supplied by the manufacture. Simulated TMI treatment planning with a slab phantom was delivered and verified with ion chamber and EDR-2 films. Results: The static output variations during 30 min averaged -2.9% ± 0.2%, -3.4% ± 0.3%, and -3.4% ± 0.3% at 10 min, 20 min, and 30 min, respectively. The rotational output variations from start averaged -2.5% ± 0.7%, -3.1% ± 0.7%, and -3.5% ± 0.8% at 10 min, 20 min, and 30 min, respectively. The maximum output variation was up to 4.5%. In a TMI planning model, in which beam-on time was over 30 min, planned dose and dose measured with ion chambers in both cranial and caudal sides agreed within 3%. Film measurements in cranial and caudal sides also showed the pass rates of 97.7% and 92.2% with the criteria of 3 mm/3% in gamma analysis. Conclusion: These results suggest that long TMI delivery by helical tomotherapy, even without dose servo system, does not pose a risk for significant deviations from the original treatment plan regardless of the output variation. However, very long time output variation should be checked before the first treatment.
AB - Background: Beam-on time in Total Marrow Irradiation (TMI) delivery with helical tomotherapy is more than 30 minutes. The purpose of this study was to investigate extended time output variation in tomotherapy machine without dose servo system and its impact on the dosimetry of TMI planning. Materials and methods: The calibration procedures with 1800 seconds delivery were conducted. The slab and cylindrical phantoms were used for static and rotational output variation measurements, respectively. All measurements were performed in 0.1 second interval with an Exradin A1SL ionization chamber (Standard Imaging Inc., Madison, WI, USA) connected to the tomoelectrometer supplied by the manufacture. Simulated TMI treatment planning with a slab phantom was delivered and verified with ion chamber and EDR-2 films. Results: The static output variations during 30 min averaged -2.9% ± 0.2%, -3.4% ± 0.3%, and -3.4% ± 0.3% at 10 min, 20 min, and 30 min, respectively. The rotational output variations from start averaged -2.5% ± 0.7%, -3.1% ± 0.7%, and -3.5% ± 0.8% at 10 min, 20 min, and 30 min, respectively. The maximum output variation was up to 4.5%. In a TMI planning model, in which beam-on time was over 30 min, planned dose and dose measured with ion chambers in both cranial and caudal sides agreed within 3%. Film measurements in cranial and caudal sides also showed the pass rates of 97.7% and 92.2% with the criteria of 3 mm/3% in gamma analysis. Conclusion: These results suggest that long TMI delivery by helical tomotherapy, even without dose servo system, does not pose a risk for significant deviations from the original treatment plan regardless of the output variation. However, very long time output variation should be checked before the first treatment.
KW - Radiation therapy
KW - Tomotherapy
KW - Total body irradiation
KW - Total marrow irradiation
KW - Very long time output variation
UR - http://www.scopus.com/inward/record.url?scp=84877853874&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=84877853874&partnerID=8YFLogxK
U2 - 10.1186/1748-717X-8-123
DO - 10.1186/1748-717X-8-123
M3 - Article
C2 - 23688298
AN - SCOPUS:84877853874
SN - 1748-717X
VL - 8
JO - Radiation Oncology
JF - Radiation Oncology
IS - 1
M1 - 123
ER -