Skip to main navigation Skip to search Skip to main content

Flow-independent dynamics in aneurysms: Intra-aneurysm pressure measurements following complete flow cessation in internal carotid artery aneurysms

  • Adnan I. Qureshi
  • , Mushtaq H. Qureshi
  • , Tanya Mohindroo
  • , Asif A. Khan
  • , Kayla Dingmann
  • , Gregory T. Sherr
  • , M. Fareed K. Suri

Research output: Contribution to journalArticlepeer-review

Abstract

Purpose: To determine if complete flow obliteration by covered stents reduces intraaneurysm pressures in internal carotid artery (ICA) aneurysms.

Methods: A single lumen microcatheter was placed into the aneurysm sac prior to covered stent deployment in 3 patients and connected to a pressure monitoring system. The intraaneurysm pressure was continuously monitored, and readings were recorded prior to and immediately after stent deployment and at 5-minute intervals up to 20 minutes after stent placement. Complete occlusion of flow into the aneurysms was confirmed by carotid angiography.

Results: There was no change in mean pressure within the aneurysm before and immediately after stent placement (80 mmHg) in any patient, nor was there a change in waveform of the intra-aneurysm pressure recording. The average of intra-aneurysm pressures among the 3 patients was higher (99 mmHg) at 10 and 15 minutes after stent placement. In 2 patients, the microcatheter was retracted into the parent arterial lumen; no difference in pressure was noted.

Conclusion: Our observations suggest no change in the pressures within the aneurysm after complete flow cessation (flow-independent). These findings may assist clinicians in better understanding aneurysm hemodynamics and rupture after covered stent deployment.

Original languageEnglish (US)
Pages (from-to)861-866
Number of pages6
JournalJournal of Endovascular Therapy
Volume21
Issue number6
DOIs
StatePublished - Dec 1 2014

Keywords

  • Aneurysm
  • Aneurysm rupture
  • Aneurysm wall
  • Compliance
  • Covered stent
  • Hemodynamics
  • Internal carotid artery
  • Intra-aneurysm pressure
  • Intrasac pressure
  • Pressure transmission
  • Stent-graft

Fingerprint

Dive into the research topics of 'Flow-independent dynamics in aneurysms: Intra-aneurysm pressure measurements following complete flow cessation in internal carotid artery aneurysms'. Together they form a unique fingerprint.

Cite this