Ferrous-ferric Ion exchange dosemeter

John A. Bauhs, Bruce E. Hammer

Research output: Contribution to journalArticlepeer-review

Abstract

In this work a three-dimensional ferrous-ferric ion exchange dosemeter is proposed and the dose response measured. The dosemeter consists of strong acid cation exchange resin beads in the H form in water. Amberlyst 15 Wet beads with a harmonic mean diameter of 0.600-0.850 mm were prepared by soaking them in an aqueous solution of ferrous ammonium sulphate to exchange ferrous ions for H + ions. The beads were rinsed with distilled water and packed in glass vials. Sets of samples with ferrous ion concentrations of 0.5 and 1.0 mM were dosed with 6 MV X rays from a Varian 2100C linac. The spin-lattice relaxation time constants (T1) for the samples were measured using an Apollo spectrometer (Tecmag, Houston, TX) interfaced to a 1.5 T magnet (Magnex, Abingdon, UK). Each sample had two T1 values; a long T1 at 1200 ms that did not significantly change with dose and a short T1 that ranged from 56 ms at 0 Gy to 36 ms at 100 Gy. The R1 vs. dose responses were linear with slopes of 0.066 and 0.079 s-1 Gy-1.

Original languageEnglish (US)
Pages (from-to)140-143
Number of pages4
JournalRadiation protection dosimetry
Volume120
Issue number1-4
DOIs
StatePublished - Sep 2006

Bibliographical note

Copyright:
Copyright 2011 Elsevier B.V., All rights reserved.

Fingerprint

Dive into the research topics of 'Ferrous-ferric Ion exchange dosemeter'. Together they form a unique fingerprint.

Cite this