Feasibility of HyperSight CBCT for adaptive radiation therapy: A phantom benchmark study of dose calculation accuracy and delivery verification on the Halcyon

  • Nicholas Nelson
  • , Courtney Oare
  • , Geoff Nelson
  • , Thomas Martin
  • , Jessica Huang
  • , Hui Zhao

Research output: Contribution to journalArticlepeer-review

Abstract

Purpose: The development of on-board cone-beam computed tomography (CBCT) has led to improved target localization and evaluation of patient anatomical change throughout the course of radiation therapy. HyperSight, a newly developed on-board CBCT platform by Varian, has been shown to improve image quality and HU fidelity relative to conventional CBCT. The purpose of this study is to benchmark the dose calculation accuracy of Varian's HyperSight cone-beam computed tomography (CBCT) on the Halcyon platform relative to fan-beam CT-based dose calculations and to perform end-to-end testing of HyperSight CBCT-only based treatment planning. Methods: A HU to mass density curve was measured for the HyperSight CBCT system and implemented into the Eclipse treatment planning system. Following this, computational dosimetric analysis was performed between dose distributions calculated on CT simulation (CTsim) and HyperSight CBCT images on two anthropomorphic phantoms for pelvic and head and neck treatment sites. Additionally, an end-to-end test was carried out for a head and neck intensity modulated radiation therapy (IMRT) plan. Results: The HU to mass density curves acquired on CTsim and HyperSight CBCT were similar (< 30 HU) for near-water equivalent materials, but deviated for high-density materials, with a maximum difference of 150 HU. For dose calculations, excellent agreement between dose calculations performed on CTsim and HyperSight CBCT phantom images was observed, where three-dimensional gamma pass rates between the two dose distributions were observed to be (Formula presented.) 90% at 1%/1 mm (5% threshold). For the end-to-end test, absolute doses were verified to within 1% of ionization chamber measurements, while Delta4+ and portal dosimetry measurements yielded passing results (gamma pass rate (Formula presented.) 90%) down to 2%/2 mm criterion. Conclusion: In this study, the accuracy of dose calculations performed on HyperSight CBCT was found to be within 1% of CTsim calculations for pelvic and head and neck treatment sites. End-to-end results using the RANDO anthropomorphic phantom indicate that HyperSight CBCT images are suitable for radiation treatment planning.

Original languageEnglish (US)
Article numbere70245
JournalJournal of Applied Clinical Medical Physics
Volume26
Issue number9
DOIs
StatePublished - Sep 2025
Externally publishedYes

Bibliographical note

Publisher Copyright:
© 2025 The Author(s). Journal of Applied Clinical Medical Physics published by Wiley Periodicals LLC on behalf of American Association of Physicists in Medicine.

Keywords

  • Halcyon
  • HyperSight
  • adaptive therapy
  • cone-beam computed tomography (CBCT)
  • dose calculations
  • dosimetry
  • photon therapy