Factor VII assay performance

An analysis of the North American Specialized Coagulation Laboratory Association proficiency testing results

N. D. Zantek, P. Hsu, M. A. Refaai, M. Ledford-Kraemer, P. Meijer, E. M. Van Cott

Research output: Contribution to journalArticle

6 Citations (Scopus)

Abstract

The performance of factor VII (FVII) assays currently used by clinical laboratories was examined in North American Specialized Coagulation Laboratory Association (NASCOLA) proficiency tests. Data from 12 surveys conducted between 2008 and 2010, involving 20 unique specimens plus four repeat-tested specimens, were analyzed. The number of laboratories per survey was 49-54 with a total of 1224 responses. Numerous reagent/instrument combinations were used. For FVII > 80 or <40 U/dL, 99.5% of results (859/863) were correctly classified by laboratories as normal/abnormal. Classification of specimens with 40-73 U/dL FVII was heterogeneous. Interlaboratory precision was better for normal specimens (coefficient of variation (CV) 10.7%) than for FVII<20 U/dL (CV 33.1%), with a mean CV of 17.2% per specimen. Intralaboratory precision for repeated specimens demonstrated no significant difference between the paired survey results (mean absolute difference 2.5-5.0 U/dL). For specimens with FVII >50 U/dL, among commonly used methods, one thromboplastin and one calibrator produced results 5-6 U/dL higher and another thromboplastin and calibrator produced results 5-6 U/dL lower than all other methods, and human thromboplastin differed from rabbit by +7.6 U/dL. Preliminary evidence suggests these differences could be due to the calibrator. For FVII <50 U/dL, differences among the commonly used reagents and calibrators were generally not significant.

Original languageEnglish (US)
Pages (from-to)314-321
Number of pages8
JournalInternational Journal of Laboratory Hematology
Volume35
Issue number3
DOIs
StatePublished - Jun 1 2013

Fingerprint

Laboratory Proficiency Testing
Factor VII
Thromboplastin
Coagulation
Assays
Testing
Clinical laboratories
Rabbits
Surveys and Questionnaires

Keywords

  • External quality control of diagnostic assays and tests
  • Factor VII
  • North American specialized coagulation laboratory association
  • Thromboplastin

Cite this

Factor VII assay performance : An analysis of the North American Specialized Coagulation Laboratory Association proficiency testing results. / Zantek, N. D.; Hsu, P.; Refaai, M. A.; Ledford-Kraemer, M.; Meijer, P.; Van Cott, E. M.

In: International Journal of Laboratory Hematology, Vol. 35, No. 3, 01.06.2013, p. 314-321.

Research output: Contribution to journalArticle

@article{777622288e31478e96acbbdb8293e9b2,
title = "Factor VII assay performance: An analysis of the North American Specialized Coagulation Laboratory Association proficiency testing results",
abstract = "The performance of factor VII (FVII) assays currently used by clinical laboratories was examined in North American Specialized Coagulation Laboratory Association (NASCOLA) proficiency tests. Data from 12 surveys conducted between 2008 and 2010, involving 20 unique specimens plus four repeat-tested specimens, were analyzed. The number of laboratories per survey was 49-54 with a total of 1224 responses. Numerous reagent/instrument combinations were used. For FVII > 80 or <40 U/dL, 99.5{\%} of results (859/863) were correctly classified by laboratories as normal/abnormal. Classification of specimens with 40-73 U/dL FVII was heterogeneous. Interlaboratory precision was better for normal specimens (coefficient of variation (CV) 10.7{\%}) than for FVII<20 U/dL (CV 33.1{\%}), with a mean CV of 17.2{\%} per specimen. Intralaboratory precision for repeated specimens demonstrated no significant difference between the paired survey results (mean absolute difference 2.5-5.0 U/dL). For specimens with FVII >50 U/dL, among commonly used methods, one thromboplastin and one calibrator produced results 5-6 U/dL higher and another thromboplastin and calibrator produced results 5-6 U/dL lower than all other methods, and human thromboplastin differed from rabbit by +7.6 U/dL. Preliminary evidence suggests these differences could be due to the calibrator. For FVII <50 U/dL, differences among the commonly used reagents and calibrators were generally not significant.",
keywords = "External quality control of diagnostic assays and tests, Factor VII, North American specialized coagulation laboratory association, Thromboplastin",
author = "Zantek, {N. D.} and P. Hsu and Refaai, {M. A.} and M. Ledford-Kraemer and P. Meijer and {Van Cott}, {E. M.}",
year = "2013",
month = "6",
day = "1",
doi = "10.1111/ijlh.12083",
language = "English (US)",
volume = "35",
pages = "314--321",
journal = "International Journal of Laboratory Hematology",
issn = "1751-5521",
publisher = "Wiley-Blackwell",
number = "3",

}

TY - JOUR

T1 - Factor VII assay performance

T2 - An analysis of the North American Specialized Coagulation Laboratory Association proficiency testing results

AU - Zantek, N. D.

AU - Hsu, P.

AU - Refaai, M. A.

AU - Ledford-Kraemer, M.

AU - Meijer, P.

AU - Van Cott, E. M.

PY - 2013/6/1

Y1 - 2013/6/1

N2 - The performance of factor VII (FVII) assays currently used by clinical laboratories was examined in North American Specialized Coagulation Laboratory Association (NASCOLA) proficiency tests. Data from 12 surveys conducted between 2008 and 2010, involving 20 unique specimens plus four repeat-tested specimens, were analyzed. The number of laboratories per survey was 49-54 with a total of 1224 responses. Numerous reagent/instrument combinations were used. For FVII > 80 or <40 U/dL, 99.5% of results (859/863) were correctly classified by laboratories as normal/abnormal. Classification of specimens with 40-73 U/dL FVII was heterogeneous. Interlaboratory precision was better for normal specimens (coefficient of variation (CV) 10.7%) than for FVII<20 U/dL (CV 33.1%), with a mean CV of 17.2% per specimen. Intralaboratory precision for repeated specimens demonstrated no significant difference between the paired survey results (mean absolute difference 2.5-5.0 U/dL). For specimens with FVII >50 U/dL, among commonly used methods, one thromboplastin and one calibrator produced results 5-6 U/dL higher and another thromboplastin and calibrator produced results 5-6 U/dL lower than all other methods, and human thromboplastin differed from rabbit by +7.6 U/dL. Preliminary evidence suggests these differences could be due to the calibrator. For FVII <50 U/dL, differences among the commonly used reagents and calibrators were generally not significant.

AB - The performance of factor VII (FVII) assays currently used by clinical laboratories was examined in North American Specialized Coagulation Laboratory Association (NASCOLA) proficiency tests. Data from 12 surveys conducted between 2008 and 2010, involving 20 unique specimens plus four repeat-tested specimens, were analyzed. The number of laboratories per survey was 49-54 with a total of 1224 responses. Numerous reagent/instrument combinations were used. For FVII > 80 or <40 U/dL, 99.5% of results (859/863) were correctly classified by laboratories as normal/abnormal. Classification of specimens with 40-73 U/dL FVII was heterogeneous. Interlaboratory precision was better for normal specimens (coefficient of variation (CV) 10.7%) than for FVII<20 U/dL (CV 33.1%), with a mean CV of 17.2% per specimen. Intralaboratory precision for repeated specimens demonstrated no significant difference between the paired survey results (mean absolute difference 2.5-5.0 U/dL). For specimens with FVII >50 U/dL, among commonly used methods, one thromboplastin and one calibrator produced results 5-6 U/dL higher and another thromboplastin and calibrator produced results 5-6 U/dL lower than all other methods, and human thromboplastin differed from rabbit by +7.6 U/dL. Preliminary evidence suggests these differences could be due to the calibrator. For FVII <50 U/dL, differences among the commonly used reagents and calibrators were generally not significant.

KW - External quality control of diagnostic assays and tests

KW - Factor VII

KW - North American specialized coagulation laboratory association

KW - Thromboplastin

UR - http://www.scopus.com/inward/record.url?scp=84876466971&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84876466971&partnerID=8YFLogxK

U2 - 10.1111/ijlh.12083

DO - 10.1111/ijlh.12083

M3 - Article

VL - 35

SP - 314

EP - 321

JO - International Journal of Laboratory Hematology

JF - International Journal of Laboratory Hematology

SN - 1751-5521

IS - 3

ER -