T1 - Experimental measurement of ligament force, facet force, and segment motion in the human lumbar spine
AU - Schendel, Michael J.
AU - Wood, Kirkham B.
AU - Buttermann, Glenn R.
AU - Lewis, Jack J.
AU - Ogilvie, James W.
PY - 1993
Y1 - 1993
N2 - Facet forces, longitudinal ligament loads, and vertebral body motion were experimentally measured in five fresh human lumbar spine segments, L1-L2. Strain gages on the bone surface were used to quantify facet loads. Buckle transducers were used to measure anterior and posterior longitudinal ligament loads. The three-dimensional motion of the motion segment was measured with an instrumented spatial linkage. The facets were found to carry no load in flexion, large loads during extension (205 N at a 10 Nm moment and a 190 N axial load), torsion (65 N at a 10 Nm moment and a 150 N axial load), and lateral bending (78 N at a 3 Nm moment and a 160 N axial load). The facet contact site on the inferior articular process of L1 was found to move inferiorly to a position of tip impingement near the lamina as extension moments increased. Impingement occurred in the range of 4-6 Nm extension. The posterior and anterior longitudinal ligaments were predominatly loaded in flexion and extension, respectively. No ligament loads occurred in lateral bending and torsion. A 1 cm strip of the anterior longitudinal ligament carried loads up to 130 N at the largest extension moment of 11.4 Nm. The posterior longitudinal ligament had a 60 N load at the largest flexion moment of 7.1 Nm. There was no pre-load in the ligament detectable with the buckle transducers (>4 N). The facets and ligaments began carrying load immediately with applied load, without a lax region. The experimental technique developed and used provides a good tool for obtaining simultaneous facet joint loads, ligament loads, and vertebral body motion without altering the motion segment.
AB - Facet forces, longitudinal ligament loads, and vertebral body motion were experimentally measured in five fresh human lumbar spine segments, L1-L2. Strain gages on the bone surface were used to quantify facet loads. Buckle transducers were used to measure anterior and posterior longitudinal ligament loads. The three-dimensional motion of the motion segment was measured with an instrumented spatial linkage. The facets were found to carry no load in flexion, large loads during extension (205 N at a 10 Nm moment and a 190 N axial load), torsion (65 N at a 10 Nm moment and a 150 N axial load), and lateral bending (78 N at a 3 Nm moment and a 160 N axial load). The facet contact site on the inferior articular process of L1 was found to move inferiorly to a position of tip impingement near the lamina as extension moments increased. Impingement occurred in the range of 4-6 Nm extension. The posterior and anterior longitudinal ligaments were predominatly loaded in flexion and extension, respectively. No ligament loads occurred in lateral bending and torsion. A 1 cm strip of the anterior longitudinal ligament carried loads up to 130 N at the largest extension moment of 11.4 Nm. The posterior longitudinal ligament had a 60 N load at the largest flexion moment of 7.1 Nm. There was no pre-load in the ligament detectable with the buckle transducers (>4 N). The facets and ligaments began carrying load immediately with applied load, without a lax region. The experimental technique developed and used provides a good tool for obtaining simultaneous facet joint loads, ligament loads, and vertebral body motion without altering the motion segment.
UR - http://www.scopus.com/inward/record.url?scp=0027577496&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=0027577496&partnerID=8YFLogxK
U2 - 10.1016/0021-9290(93)90006-Z
DO - 10.1016/0021-9290(93)90006-Z
M3 - Article
C2 - 8478347
AN - SCOPUS:0027577496
SN - 0021-9290
VL - 26
SP - 427
EP - 438
JO - Journal of Biomechanics
JF - Journal of Biomechanics
IS - 4-5
ER -