Experimental determination of the BOLD field strength dependence in vessels and tissue

Joseph S. Gati, Ravi S. Menon, Kâmil Uǧurbil, Brian K. Rutt

Research output: Contribution to journalArticlepeer-review

424 Scopus citations

Abstract

High resolution functional MRI (fMRI) experiments were performed in human visual cortex at 0.5, 1.5, and 4 T to determine the blood oxygenation level dependent (BOLD) field strength response within regions of obvious venous vessels and cortical gray matter ('tissue'). T2*-weighted FLASH images were collected in single- and multi-echo mode and used to determine the intrinsio BOLD parameters, namely, signal-to-noise ratio (Ψ), the apparent transverse relaxation rate (R2*) and the change in R2* (ΔR2*) between the activated and baseline states. The authors find the average percentage signal change (ΔS/S, measured at TE = T2*) to be large in vessels (13.3 ± 2.3%, 18.4 ± 4.0%, and 15.1 ± 1.2%) compared with that in tissue (1.4 ± 0.7%, 1.9 ± 0.7%, and 3.3 ± 0.2%) at 0.5, 1.5, and 4 T, respectively. The signal-to-noise ratio in optimized, fully relaxed proton density weighted gradient echo images was found to increase linearly with respect to the static magnetic field strength (B0). The predicted upper bound on BOLD contrast-to-noise ratio (ΔS/R)(max) as a function of field strength was calculated and found to behave less than linearly in voxels containing vessels larger than the voxel itself and greater than linearly in voxels containing a mixture of capillaries and veins/venules with a diameter less than that of the voxel.

Original languageEnglish (US)
Pages (from-to)296-302
Number of pages7
JournalMagnetic resonance in medicine
Volume38
Issue number2
DOIs
StatePublished - Aug 1997

Keywords

  • BOLD
  • Cortical gray matter
  • Field strength
  • Venous vessels

Fingerprint

Dive into the research topics of 'Experimental determination of the BOLD field strength dependence in vessels and tissue'. Together they form a unique fingerprint.

Cite this