Electrical parameters for physiological His–Purkinje pacing vary by implant location in an ex vivo canine model

Alexander R. Mattson, Elizabeth Mattson, Mary Lauren Mesich, Zhongping Yang, Paul A. Iaizzo

Research output: Contribution to journalArticlepeer-review

7 Scopus citations

Abstract

Background: Permanent His-bundle pacing (HBP) is an attractive, perhaps more physiological, alternative to traditional right ventricular pacing. Objective: The purpose of this study was to utilize direct visualization to more comprehensively understand the anatomy central to HBP, correlating electrical lead performance to implant locations along the His-bundle (HB) pathway. Methods: Canine hearts (n = 5) were isolated and reanimated using Visible Heart methodologies. Medtronic 3830 SelectSecure leads were fixated where His potentials were present. The location of each implant was mapped/binned into 4 regions approximately analogous to the proximal, penetrating, and distal HB. Locational differences in HBP capture and resultant QRS morphology were assessed. Results: Average HBP capture thresholds did not significantly vary with respect to implant location (1.0-ms pulse width; P =.48). The resulting QRS morphologies from HB-paced beats varied in relation to implant location. As leads were placed further distally along the HB, the ratio of paced to native QRS complex duration increased (ΔQRS paced /ΔQRS native ratios—region 2: 0.84 ± 0.16; region 3: 1.04 ± 0.42; region 4: 1.74 ± 0.86). Conclusion: We demonstrated correlation between the anatomic locations of HBP lead placement and resultant QRS morphologies in a reanimated canine heart model. Proximal placement along the HB pathway resulted in more favorable QRS morphologies, suggesting improved selective HBP capture, with no significant increase in HBP capture thresholds. Pacing the HB in more proximal pathway locations improved the selectivity of HBP and may confer electrical and anatomic benefits relative to distal HBP.

Original languageEnglish (US)
Pages (from-to)443-450
Number of pages8
JournalHeart Rhythm
Volume16
Issue number3
DOIs
StatePublished - Mar 2019

Bibliographical note

Funding Information:
Sources of funding: This work was supported by a research contract with Medtronic . The sponsor did not have a role in study design, data collection/analysis/interpretation, report writing, or decision to submit for publication; Institute for Engineering in Medicine, University of Minnesota. Disclosures: Dr A.R. Mattson is employed and has stock ownership with Medtronic. Ms E. Mattson is employed by Medtronic. Drs Mesich and Yang are employed by and have stock ownership with Medtronic. Dr Iaizzo has a research contract with Medtronic.

Publisher Copyright:
© 2018 Heart Rhythm Society

Keywords

  • Direct visualization
  • His bundle
  • His–Purkinje system
  • Membranous septum
  • Pacing