Development of an Integrated C-Arm Interventional Imaging System with a Strip Photon Counting Detector and a Flat Panel Detector

  • Xu Ji
  • , Mang Feng
  • , Kevin Treb
  • , Ran Zhang
  • , Sebastian Schafer
  • , Ke Li

Research output: Contribution to journalArticlepeer-review

Abstract

Modern interventional x-ray systems are often equipped with flat-panel detector-based cone-beam CT (FPD-CBCT) to provide tomographic, volumetric, and high spatial resolution imaging of interventional devices, iodinated vessels, and other objects. The purpose of this work was to bring an interchangeable strip photon-counting detector (PCD) to C-arm systems to supplement (instead of retiring) the existing FPD-CBCT with a high quality, spectral, and affordable PCD-CT imaging option. With minimal modification to the existing C-arm, a 51 × 0.6 cm2 PCD with a 0.75 mm CdTe layer, two energy thresholds, and 0.1 mm pixels was integrated with a Siemens Artis Zee interventional imaging system. The PCD can be translated in and out of the field-of-view to allow the system to switch between FPD and PCD-CT imaging modes. A dedicated phantom and a new algorithm were developed to calibrate the projection geometry of the narrow-beam PCD-CT system and correct the gantry wobbling-induced geometric distortion artifacts. In addition, a detector response calibration procedure was performed for each PCD pixel using materials with known radiological pathlengths to address concentric artifacts in PCD-CT images. Both phantom and human cadaver experiments were performed at a high gantry rotation speed and clinically relevant radiation dose level to evaluate the spectral and non-spectral imaging performance of the prototype system. Results show that the PCD-CT system has excellent image quality with negligible artifacts after the proposed corrections. Compared with FPD-CBCT images acquired at the same dose level, PCD-CT images demonstrated a 53% reduction in noise variance and additional quantitative imaging capability.

Original languageEnglish (US)
Pages (from-to)3674-3685
Number of pages12
JournalIEEE Transactions on Medical Imaging
Volume40
Issue number12
DOIs
StatePublished - Dec 1 2021
Externally publishedYes

Bibliographical note

Publisher Copyright:
© 1982-2012 IEEE.

Keywords

  • C-arm x-ray systems
  • image guided intervention
  • photon counting CT
  • Photon counting detector