Design and evaluation of an RF front-end for 9.4 T human MRI

G. Shajan, Jens Hoffmann, Juliane Budde, Gregor Adriany, Kamil Ugurbil, Rolf Pohmann

Research output: Contribution to journalArticle

18 Citations (Scopus)

Abstract

At the field strength of 9.4 T, the highest field currently available for human MRI, the wavelength of the MR signals is significantly shorter than the size of the examined structures. Even more than at 7 T, constructive and destructive interferences cause strong inhomogeneities of the B 1 field produced by a volume coil, causing shading over large parts of the image. Specialized radio frequency hardware and B 1 management methods are required to obtain high-quality images that take full advantage of the high field strength. Here, the design and characteristics of a radio frequency front-end especially developed for proton imaging at 9.4 T are presented. In addition to a 16-channel transceiver array coil, capable of volume transmit mode and independent signal reception, it consists of custom built low noise preamplifiers and TR switches. Destructive interference patterns were eliminated, in virtually the entire brain, using a simple in situ radio frequency phase shimming technique. After mapping the Ba 1 + profile of each transmit channel, a numerical algorithm was used to calculate the appropriate transmit phase offsets needed to obtain a homogeneous excitation field over a user defined region. Between two and three phase settings are necessary to obtain homogeneous images over the entire brain.

Original languageEnglish (US)
Pages (from-to)596-604
Number of pages9
JournalMagnetic resonance in medicine
Volume66
Issue number2
DOIs
StatePublished - Aug 1 2011

Fingerprint

Radio
Brain
Noise
Protons

Keywords

  • 9.4 T human brain MRI
  • B shimming
  • microstrip transmission line
  • transceiver array

Cite this

Design and evaluation of an RF front-end for 9.4 T human MRI. / Shajan, G.; Hoffmann, Jens; Budde, Juliane; Adriany, Gregor; Ugurbil, Kamil; Pohmann, Rolf.

In: Magnetic resonance in medicine, Vol. 66, No. 2, 01.08.2011, p. 596-604.

Research output: Contribution to journalArticle

Shajan, G. ; Hoffmann, Jens ; Budde, Juliane ; Adriany, Gregor ; Ugurbil, Kamil ; Pohmann, Rolf. / Design and evaluation of an RF front-end for 9.4 T human MRI. In: Magnetic resonance in medicine. 2011 ; Vol. 66, No. 2. pp. 596-604.
@article{90fef5f97be541e7840e6a67ae79bcea,
title = "Design and evaluation of an RF front-end for 9.4 T human MRI",
abstract = "At the field strength of 9.4 T, the highest field currently available for human MRI, the wavelength of the MR signals is significantly shorter than the size of the examined structures. Even more than at 7 T, constructive and destructive interferences cause strong inhomogeneities of the B 1 field produced by a volume coil, causing shading over large parts of the image. Specialized radio frequency hardware and B 1 management methods are required to obtain high-quality images that take full advantage of the high field strength. Here, the design and characteristics of a radio frequency front-end especially developed for proton imaging at 9.4 T are presented. In addition to a 16-channel transceiver array coil, capable of volume transmit mode and independent signal reception, it consists of custom built low noise preamplifiers and TR switches. Destructive interference patterns were eliminated, in virtually the entire brain, using a simple in situ radio frequency phase shimming technique. After mapping the Ba 1 + profile of each transmit channel, a numerical algorithm was used to calculate the appropriate transmit phase offsets needed to obtain a homogeneous excitation field over a user defined region. Between two and three phase settings are necessary to obtain homogeneous images over the entire brain.",
keywords = "9.4 T human brain MRI, B shimming, microstrip transmission line, transceiver array",
author = "G. Shajan and Jens Hoffmann and Juliane Budde and Gregor Adriany and Kamil Ugurbil and Rolf Pohmann",
year = "2011",
month = "8",
day = "1",
doi = "10.1002/mrm.22808",
language = "English (US)",
volume = "66",
pages = "596--604",
journal = "Magnetic Resonance in Medicine",
issn = "0740-3194",
publisher = "John Wiley and Sons Inc.",
number = "2",

}

TY - JOUR

T1 - Design and evaluation of an RF front-end for 9.4 T human MRI

AU - Shajan, G.

AU - Hoffmann, Jens

AU - Budde, Juliane

AU - Adriany, Gregor

AU - Ugurbil, Kamil

AU - Pohmann, Rolf

PY - 2011/8/1

Y1 - 2011/8/1

N2 - At the field strength of 9.4 T, the highest field currently available for human MRI, the wavelength of the MR signals is significantly shorter than the size of the examined structures. Even more than at 7 T, constructive and destructive interferences cause strong inhomogeneities of the B 1 field produced by a volume coil, causing shading over large parts of the image. Specialized radio frequency hardware and B 1 management methods are required to obtain high-quality images that take full advantage of the high field strength. Here, the design and characteristics of a radio frequency front-end especially developed for proton imaging at 9.4 T are presented. In addition to a 16-channel transceiver array coil, capable of volume transmit mode and independent signal reception, it consists of custom built low noise preamplifiers and TR switches. Destructive interference patterns were eliminated, in virtually the entire brain, using a simple in situ radio frequency phase shimming technique. After mapping the Ba 1 + profile of each transmit channel, a numerical algorithm was used to calculate the appropriate transmit phase offsets needed to obtain a homogeneous excitation field over a user defined region. Between two and three phase settings are necessary to obtain homogeneous images over the entire brain.

AB - At the field strength of 9.4 T, the highest field currently available for human MRI, the wavelength of the MR signals is significantly shorter than the size of the examined structures. Even more than at 7 T, constructive and destructive interferences cause strong inhomogeneities of the B 1 field produced by a volume coil, causing shading over large parts of the image. Specialized radio frequency hardware and B 1 management methods are required to obtain high-quality images that take full advantage of the high field strength. Here, the design and characteristics of a radio frequency front-end especially developed for proton imaging at 9.4 T are presented. In addition to a 16-channel transceiver array coil, capable of volume transmit mode and independent signal reception, it consists of custom built low noise preamplifiers and TR switches. Destructive interference patterns were eliminated, in virtually the entire brain, using a simple in situ radio frequency phase shimming technique. After mapping the Ba 1 + profile of each transmit channel, a numerical algorithm was used to calculate the appropriate transmit phase offsets needed to obtain a homogeneous excitation field over a user defined region. Between two and three phase settings are necessary to obtain homogeneous images over the entire brain.

KW - 9.4 T human brain MRI

KW - B shimming

KW - microstrip transmission line

KW - transceiver array

UR - http://www.scopus.com/inward/record.url?scp=79960572853&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=79960572853&partnerID=8YFLogxK

U2 - 10.1002/mrm.22808

DO - 10.1002/mrm.22808

M3 - Article

C2 - 21381102

AN - SCOPUS:79960572853

VL - 66

SP - 596

EP - 604

JO - Magnetic Resonance in Medicine

JF - Magnetic Resonance in Medicine

SN - 0740-3194

IS - 2

ER -