Abstract
A channel plate multiplier was placed between the specimen and scintillator to increase the number of secondary electrons in the emissive mode of operation of the SEM. A comparison between this system and the conventional scintillator detector shows that the presence of the plate improves the signal-to-noise ratio.
Original language | English (US) |
---|---|
Pages (from-to) | 806-808 |
Number of pages | 3 |
Journal | Review of Scientific Instruments |
Volume | 48 |
Issue number | 7 |
DOIs | |
State | Published - 1977 |