Autonomic regulation therapy via left or right cervical vagus nerve stimulation in patients with chronic heart failure: Results of the ANTHEM-HF trial

Rajendra K. Premchand, Kamal Sharma, Sanjay Mittal, Rufino Monteiro, Satyajit Dixit, Imad Libbus, Lorenzo A. Dicarlo, Jeffrey L. Ardell, Thomas S. Rector, Badri Amurthur, Bruce H. Kenknight, Inder S. Anand

Research output: Contribution to journalArticlepeer-review

295 Scopus citations

Abstract

Objective ANTHEM-HF evaluated a novel autonomic regulation therapy (ART) via either left or right vagus nerve stimulation (VNS) in patients with heart failure (HF) and reduced ejection fraction (HFrEF).

Methods and Results Sixty subjects (New York Heart Association [NYHA] functional class II-III, left ventricular ejection fraction (LVEF) ≤40%, left ventricular end-diastolic diameter ≥50 mm to ≤80 mm) receiving optimal pharmacologic therapy were randomized at 10 sites. VNS systems were randomly implanted on the left (n = 31) or right (n = 29) side. All patients were successfully implanted and 59 were titrated over 10 weeks to a well tolerated stimulation intensity. One patient died 3 days after an embolic stroke that occurred during implantation. Common device-related adverse events after VNS titration were transient mild dysphonia, cough, and oropharyngeal pain, which were similar for left- and right-side VNS. After 6 months of ART, the adjusted left-right differences in LVEF, left ventricular end-systolic volume (LVESV), and left ventricular end-systolic diameter (LVESD) were 0.2% (95% CI -4.4 to 4.7), 3.7 mL (95% CI -7.0 to 14.4), and 1.3 mm (95% CI -0.9 to 3.6), respectively. In the combined population, absolute LVEF improved by 4.5% (95% CI 2.4-6.6), LVESV improved by -4.1 mL (95% CI -9.0 to 0.8), and LVESD improved by -1.7 mm (95% CI -2.8 to -0.7). Heart rate variability improved by 17 ms (95% CI 6.5-28) with minimal left-right difference. Six-minute walk distance improved an average of 56 m (95% CI 37-75); however, improvement was greater for right-side ART (77 m [95% CI 49-105]). NYHA functional class improved in 77% of patients (baseline to 6 months).

Conclusions Chronic open-loop ART via left- or right-side VNS is feasible and well tolerated in HFrEF patients. Safety and efficacy measures are encouraging and warrant further study.

Original languageEnglish (US)
Pages (from-to)808-816
Number of pages9
JournalJournal of cardiac failure
Volume20
Issue number11
DOIs
StatePublished - Nov 1 2014

Bibliographical note

Publisher Copyright:
© 2014 Elsevier Inc.