TY - JOUR
T1 - Autonomic regulation therapy via left or right cervical vagus nerve stimulation in patients with chronic heart failure
T2 - Results of the ANTHEM-HF trial
AU - Premchand, Rajendra K.
AU - Sharma, Kamal
AU - Mittal, Sanjay
AU - Monteiro, Rufino
AU - Dixit, Satyajit
AU - Libbus, Imad
AU - Dicarlo, Lorenzo A.
AU - Ardell, Jeffrey L.
AU - Rector, Thomas S.
AU - Amurthur, Badri
AU - Kenknight, Bruce H.
AU - Anand, Inder S.
N1 - Publisher Copyright:
© 2014 Elsevier Inc.
PY - 2014/11/1
Y1 - 2014/11/1
N2 - Objective ANTHEM-HF evaluated a novel autonomic regulation therapy (ART) via either left or right vagus nerve stimulation (VNS) in patients with heart failure (HF) and reduced ejection fraction (HFrEF).Methods and Results Sixty subjects (New York Heart Association [NYHA] functional class II-III, left ventricular ejection fraction (LVEF) ≤40%, left ventricular end-diastolic diameter ≥50 mm to ≤80 mm) receiving optimal pharmacologic therapy were randomized at 10 sites. VNS systems were randomly implanted on the left (n = 31) or right (n = 29) side. All patients were successfully implanted and 59 were titrated over 10 weeks to a well tolerated stimulation intensity. One patient died 3 days after an embolic stroke that occurred during implantation. Common device-related adverse events after VNS titration were transient mild dysphonia, cough, and oropharyngeal pain, which were similar for left- and right-side VNS. After 6 months of ART, the adjusted left-right differences in LVEF, left ventricular end-systolic volume (LVESV), and left ventricular end-systolic diameter (LVESD) were 0.2% (95% CI -4.4 to 4.7), 3.7 mL (95% CI -7.0 to 14.4), and 1.3 mm (95% CI -0.9 to 3.6), respectively. In the combined population, absolute LVEF improved by 4.5% (95% CI 2.4-6.6), LVESV improved by -4.1 mL (95% CI -9.0 to 0.8), and LVESD improved by -1.7 mm (95% CI -2.8 to -0.7). Heart rate variability improved by 17 ms (95% CI 6.5-28) with minimal left-right difference. Six-minute walk distance improved an average of 56 m (95% CI 37-75); however, improvement was greater for right-side ART (77 m [95% CI 49-105]). NYHA functional class improved in 77% of patients (baseline to 6 months).Conclusions Chronic open-loop ART via left- or right-side VNS is feasible and well tolerated in HFrEF patients. Safety and efficacy measures are encouraging and warrant further study.
AB - Objective ANTHEM-HF evaluated a novel autonomic regulation therapy (ART) via either left or right vagus nerve stimulation (VNS) in patients with heart failure (HF) and reduced ejection fraction (HFrEF).Methods and Results Sixty subjects (New York Heart Association [NYHA] functional class II-III, left ventricular ejection fraction (LVEF) ≤40%, left ventricular end-diastolic diameter ≥50 mm to ≤80 mm) receiving optimal pharmacologic therapy were randomized at 10 sites. VNS systems were randomly implanted on the left (n = 31) or right (n = 29) side. All patients were successfully implanted and 59 were titrated over 10 weeks to a well tolerated stimulation intensity. One patient died 3 days after an embolic stroke that occurred during implantation. Common device-related adverse events after VNS titration were transient mild dysphonia, cough, and oropharyngeal pain, which were similar for left- and right-side VNS. After 6 months of ART, the adjusted left-right differences in LVEF, left ventricular end-systolic volume (LVESV), and left ventricular end-systolic diameter (LVESD) were 0.2% (95% CI -4.4 to 4.7), 3.7 mL (95% CI -7.0 to 14.4), and 1.3 mm (95% CI -0.9 to 3.6), respectively. In the combined population, absolute LVEF improved by 4.5% (95% CI 2.4-6.6), LVESV improved by -4.1 mL (95% CI -9.0 to 0.8), and LVESD improved by -1.7 mm (95% CI -2.8 to -0.7). Heart rate variability improved by 17 ms (95% CI 6.5-28) with minimal left-right difference. Six-minute walk distance improved an average of 56 m (95% CI 37-75); however, improvement was greater for right-side ART (77 m [95% CI 49-105]). NYHA functional class improved in 77% of patients (baseline to 6 months).Conclusions Chronic open-loop ART via left- or right-side VNS is feasible and well tolerated in HFrEF patients. Safety and efficacy measures are encouraging and warrant further study.
KW - Heart failure
KW - autonomic regulation therapy
KW - nonpharmacologic therapy
KW - vagus nerve stimulation
UR - http://www.scopus.com/inward/record.url?scp=84908373096&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=84908373096&partnerID=8YFLogxK
U2 - 10.1016/j.cardfail.2014.08.009
DO - 10.1016/j.cardfail.2014.08.009
M3 - Article
C2 - 25187002
AN - SCOPUS:84908373096
SN - 1071-9164
VL - 20
SP - 808
EP - 816
JO - Journal of cardiac failure
JF - Journal of cardiac failure
IS - 11