Skip to main navigation Skip to search Skip to main content

Amperometric Biosensors Based on a Biocatalyst Electrode with Entrapped Mediator

  • Mitsugi Senda
  • , Tokuji Ikeda
  • , Kojiro Miki
  • , Hiroshi Hiasa

Research output: Contribution to journalArticlepeer-review

Abstract

A mathematical model of a biocatalyst electrode with entrapped mediator is presented. The electrode is a benzoquinone (BQ)-mixed carbon paste electrode (CPE), on which glucose oxidase (GOD) is immobilized by coating with a polymer film or semipermeable membrane (s.m., a film-coated GOD-BQ-CPE). BQ is dissolved in the immobilized-enzyme layer(e.L) to mediate the electron transfer between the electrode and the enzyme. The model is focused on the diffusion accompanied with enzyme reaction of substrate (D-glucose) and mediators (BQ and its reduced form) in the e.l. and the diffusion in the s.m. The amperometric response of the electrode is analyzed as a function of transport, kinetic, and geometrical parameters and compared with experimental results. A glucose sensor based on a film-coated GOD-BQ-CPE has a wide concentration range of response and is not affected by oxygen in test solution. Biosensors for D-gluconate and ethanol based on film-coated gluconate dehydrogenose-BQ- and alcohol dehydrogenase-NAD-CPE's, respectively, are presented.

Original languageEnglish (US)
Pages (from-to)501-506
Number of pages6
JournalAnalytical Sciences
Volume2
Issue number6
DOIs
StatePublished - 1986

Keywords

  • Biocatalyst electrode
  • biosensor
  • ethanol sensor
  • gluconate sensor
  • glucose sensor

Fingerprint

Dive into the research topics of 'Amperometric Biosensors Based on a Biocatalyst Electrode with Entrapped Mediator'. Together they form a unique fingerprint.

Cite this