The aim of our study was to investigate the accuracy of targeting using intraoperative "O-arm" during deep brain stimulation (DBS) surgery. Intraoperative O-arm (Medtronic, Minneapolis, MN, USA) images were obtained to confirm the accuracy of placement. The difference between intended and actual target coordinates was calculated based on intraoperative images and postoperative CT scan. Euclidian vector error was obtained to estimate the directional error. Correlation of targeting error with the pneumocephalus and the deviation from the planned trajectory was also estimated. Twenty eight DBS leads (globus pallidus internus [GPi], n = 13; subthalamic nucleus [STN], n = 9; ventralis intermedius nucleus [VIM], n = 6) were implanted in 20 patients using the stereotactic Leksell frame (Elekta AB, Stockholm, Sweden) under general anesthesia over a period of 1 year. The mean age was 63.6 ± standard error of the mean (SEM) 15.7 years and 60% of patients were males. The mean absolute difference (+SEM) between intended and actual target in x, y and z coordinates based on intraoperative CT scan was 0.65 ± 0.09 (p = 0.84), 0.58 ± 0.08 (p = 0.98), 1.13 ± 0.10 (p = 0.08), respectively, and postoperative (1 month) CT scan was 0.82 ± 0.15 (p = 0.89), 0.55 ± 0.11 (p = 0.97), and 1.58 ± 0.29 (p = 0.08), respectively. The Euclidean vector error was 1.59 ± 0.10 and 2.16 ± 0.26 based on intraoperative and postoperative images, respectively. There was no statistically significant targeting error based on fusion of intraoperative CT images to either preoperative CT scan or MRI as registration series, the presence of pneumocephalus, deviation from planned trajectory or the anatomical target (STN versus VIM versus GPi) (p > 0.05). Superficial skin infection was encountered in a single patient in this study. The mean total operating room time was 193.5 ± 74.6 minutes. None of the patients required revision in our study. DBS leads can be implanted safely and accurately using intraoperative O-arm with a frame based targeting system and can be used in selected patients.

Original languageEnglish (US)
Pages (from-to)80-86
Number of pages7
JournalJournal of Clinical Neuroscience
Volume27
DOIs
StatePublished - May 1 2016
Externally publishedYes

Bibliographical note

Publisher Copyright:
© 2015 Elsevier Ltd.

Keywords

  • Accuracy
  • Deep brain stimulation
  • Image guidance
  • Intraoperative
  • Neuromodulation
  • Safety

Fingerprint

Dive into the research topics of 'Accuracy and safety of targeting using intraoperative "o-arm" during placement of deep brain stimulation electrodes without electrophysiological recordings'. Together they form a unique fingerprint.

Cite this