Abstract
A new and simple technique has been developed using plastic scintillator detectors for gamma ray and cosmic ray spectroscopy without single channel analyzer (SCA) or multichannel analyzer (MCA). In these experiments only a leading edge discriminator (LED) and NIM scalers have been used. Energy calibration of gamma spectra in plastic scintillators has been done using Co60 and Cs137 sources. The details of the experimental technique, analysis procedure and experimental results have been presented in this paper.
Original language | English (US) |
---|---|
Pages (from-to) | 606-608 |
Number of pages | 3 |
Journal | Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment |
Volume | 824 |
DOIs | |
State | Published - Jul 11 2016 |
Externally published | Yes |
Bibliographical note
Publisher Copyright:© 2015 Elsevier B.V. All rights reserved.
Keywords
- Cosmic ray
- Energy resolution
- Gamma ray
- Plastic scintillator
- Spectroscopy