A review of state-of-the-art numerical methods for simulating flow through mechanical heart valves

Fotis Sotiropoulos, Iman Borazjani

Research output: Contribution to journalReview article

72 Citations (Scopus)

Abstract

In nearly half of the heart valve replacement surgeries performed annually, surgeons prefer to implant bileaflet mechanical heart valves (BMHV) because of their durability and long life span. All current BMHV designs, however, are prone to thromboembolic complications and implant recipients need to be on a life-long anticoagulant medication regiment. Non-physiologic flow patterns and turbulence generated by the valve leaflets are believed to be the major culprit for the increased risk of thromboembolism in BMHV implant recipients. In this paper, we review recent advances in developing predictive fluid-structure interaction (FSI) algorithms that can simulate BMHV flows at physiologic conditions and at resolution sufficiently fine to start probing the links between hemodynamics and blood-cell damage. Numerical simulations have provided the first glimpse into the complex hemodynamic environment experienced by blood cells downstream of the valve leaflets and successfully resolved for the first time the experimentally observed explosive transition to a turbulent-like state at the start of the decelerating flow phase. The simulations have also resolved a number of subtle features of experimentally observed valve kinematics, such as the asymmetric opening and closing of the leaflets and the leaflet rebound during closing. The paper also discusses a future research agenda toward developing a powerful patient-specific computational framework for optimizing valve design and implantation in a virtual surgery environment.

Original languageEnglish (US)
Pages (from-to)245-256
Number of pages12
JournalMedical and Biological Engineering and Computing
Volume47
Issue number3
DOIs
StatePublished - Feb 12 2009

Fingerprint

Numerical methods
Hemodynamics
Surgery
Blood
Cells
Fluid structure interaction
Flow patterns
Kinematics
Durability
Turbulence
Computer simulation

Keywords

  • Aorta
  • Immersed boundary
  • Mechanical heart valve
  • Numerical method
  • Simulation

Cite this

A review of state-of-the-art numerical methods for simulating flow through mechanical heart valves. / Sotiropoulos, Fotis; Borazjani, Iman.

In: Medical and Biological Engineering and Computing, Vol. 47, No. 3, 12.02.2009, p. 245-256.

Research output: Contribution to journalReview article

@article{9aae68becd0040f3915a8c37a9a068fc,
title = "A review of state-of-the-art numerical methods for simulating flow through mechanical heart valves",
abstract = "In nearly half of the heart valve replacement surgeries performed annually, surgeons prefer to implant bileaflet mechanical heart valves (BMHV) because of their durability and long life span. All current BMHV designs, however, are prone to thromboembolic complications and implant recipients need to be on a life-long anticoagulant medication regiment. Non-physiologic flow patterns and turbulence generated by the valve leaflets are believed to be the major culprit for the increased risk of thromboembolism in BMHV implant recipients. In this paper, we review recent advances in developing predictive fluid-structure interaction (FSI) algorithms that can simulate BMHV flows at physiologic conditions and at resolution sufficiently fine to start probing the links between hemodynamics and blood-cell damage. Numerical simulations have provided the first glimpse into the complex hemodynamic environment experienced by blood cells downstream of the valve leaflets and successfully resolved for the first time the experimentally observed explosive transition to a turbulent-like state at the start of the decelerating flow phase. The simulations have also resolved a number of subtle features of experimentally observed valve kinematics, such as the asymmetric opening and closing of the leaflets and the leaflet rebound during closing. The paper also discusses a future research agenda toward developing a powerful patient-specific computational framework for optimizing valve design and implantation in a virtual surgery environment.",
keywords = "Aorta, Immersed boundary, Mechanical heart valve, Numerical method, Simulation",
author = "Fotis Sotiropoulos and Iman Borazjani",
year = "2009",
month = "2",
day = "12",
doi = "10.1007/s11517-009-0438-z",
language = "English (US)",
volume = "47",
pages = "245--256",
journal = "Medical and biological engineering",
issn = "0140-0118",
publisher = "Peter Peregrinus, Ltd.",
number = "3",

}

TY - JOUR

T1 - A review of state-of-the-art numerical methods for simulating flow through mechanical heart valves

AU - Sotiropoulos, Fotis

AU - Borazjani, Iman

PY - 2009/2/12

Y1 - 2009/2/12

N2 - In nearly half of the heart valve replacement surgeries performed annually, surgeons prefer to implant bileaflet mechanical heart valves (BMHV) because of their durability and long life span. All current BMHV designs, however, are prone to thromboembolic complications and implant recipients need to be on a life-long anticoagulant medication regiment. Non-physiologic flow patterns and turbulence generated by the valve leaflets are believed to be the major culprit for the increased risk of thromboembolism in BMHV implant recipients. In this paper, we review recent advances in developing predictive fluid-structure interaction (FSI) algorithms that can simulate BMHV flows at physiologic conditions and at resolution sufficiently fine to start probing the links between hemodynamics and blood-cell damage. Numerical simulations have provided the first glimpse into the complex hemodynamic environment experienced by blood cells downstream of the valve leaflets and successfully resolved for the first time the experimentally observed explosive transition to a turbulent-like state at the start of the decelerating flow phase. The simulations have also resolved a number of subtle features of experimentally observed valve kinematics, such as the asymmetric opening and closing of the leaflets and the leaflet rebound during closing. The paper also discusses a future research agenda toward developing a powerful patient-specific computational framework for optimizing valve design and implantation in a virtual surgery environment.

AB - In nearly half of the heart valve replacement surgeries performed annually, surgeons prefer to implant bileaflet mechanical heart valves (BMHV) because of their durability and long life span. All current BMHV designs, however, are prone to thromboembolic complications and implant recipients need to be on a life-long anticoagulant medication regiment. Non-physiologic flow patterns and turbulence generated by the valve leaflets are believed to be the major culprit for the increased risk of thromboembolism in BMHV implant recipients. In this paper, we review recent advances in developing predictive fluid-structure interaction (FSI) algorithms that can simulate BMHV flows at physiologic conditions and at resolution sufficiently fine to start probing the links between hemodynamics and blood-cell damage. Numerical simulations have provided the first glimpse into the complex hemodynamic environment experienced by blood cells downstream of the valve leaflets and successfully resolved for the first time the experimentally observed explosive transition to a turbulent-like state at the start of the decelerating flow phase. The simulations have also resolved a number of subtle features of experimentally observed valve kinematics, such as the asymmetric opening and closing of the leaflets and the leaflet rebound during closing. The paper also discusses a future research agenda toward developing a powerful patient-specific computational framework for optimizing valve design and implantation in a virtual surgery environment.

KW - Aorta

KW - Immersed boundary

KW - Mechanical heart valve

KW - Numerical method

KW - Simulation

UR - http://www.scopus.com/inward/record.url?scp=62149095772&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=62149095772&partnerID=8YFLogxK

U2 - 10.1007/s11517-009-0438-z

DO - 10.1007/s11517-009-0438-z

M3 - Review article

VL - 47

SP - 245

EP - 256

JO - Medical and biological engineering

JF - Medical and biological engineering

SN - 0140-0118

IS - 3

ER -