Abstract
Accurate mass measurements of 20 drugs were conducted using MassWorks software on a TSQ Quantum Ultra mass spectrometer. All Q1 scan mass spectral data were collected in profile mode with full width at half-maximum (FWHM) set at 0. 7 Da. The mass errors of all 20 drugs (molecular weight between 135 and 810) were within the range from - 22. 4 × 10-6 to 36. 1 × 10-6, among them 90% of the drugs achieved a mass accuracy better than 10 × 10-6. The new method can provide accurate mass measurements on unit mass resolution mass spectrometers.
| Original language | English (US) |
|---|---|
| Pages (from-to) | 1112-1114 |
| Number of pages | 3 |
| Journal | Yaoxue Xuebao |
| Volume | 42 |
| Issue number | 10 |
| State | Published - Oct 1 2007 |
Keywords
- Accurate relative molecular mass
- Isotope
- Unit mass resolution mass spectrometer
Fingerprint
Dive into the research topics of 'A new software method for accurate mass measurements of drugs on unit mass resolution mass spectrometer'. Together they form a unique fingerprint.Cite this