A Model to Evaluate Alternative Methods of Defibrillation Threshold Determination

TIMOTHY CHURCH, MELISSA MARTINSON, MICHAEL KALLOK, WARREN WATSON

Research output: Contribution to journalArticlepeer-review

26 Scopus citations

Abstract

The voltage (or, equivalently, energy) at which defibrillation occurs for a specific episode of fibrillation can be represented by specifying or estimating the probability of successful defibrillation for each voltage or energy. This relation of voltage to probability is the probability function. For a series of attempts, the probability function predicts the frequency with which defibrillation will occur at a given voltage. By defining the defibrillation threshold (DFT) as the voltage at which the probability function takes on a specific value, say 50%, a method of defibrillation threshold determination can be evaluated by how accurately and precisely it estimates the “true” defibrillation threshold. By utilizing estimated probability functions from animals and humans, the relative performance of different methods of defibrillation threshold determination can be evaluated. Three methods were evaluated using published animal data and human clinical data: (1) stepping down to the first voltage that fails; (2) stepping up to the first voltage that succeeds; and (3) doing both 1 and 2 and averaging. In all cases, Method 3 has lower total error. Definitions of defibrillation threshold other than the 50% level can also be evaluated in this fashion.

Original languageEnglish (US)
Pages (from-to)2002-2007
Number of pages6
JournalPacing and Clinical Electrophysiology
Volume11
Issue number11
DOIs
StatePublished - Nov 1988
Externally publishedYes

Keywords

  • defibrillation threshold
  • ventricular fibrillation

Fingerprint

Dive into the research topics of 'A Model to Evaluate Alternative Methods of Defibrillation Threshold Determination'. Together they form a unique fingerprint.

Cite this