Abstract
A new adiabatic pulse, which can induce uniform and arbitrary flip angles despite the presence of transmitter coil magnetic field (B1) inhomogeneities, is employed for 3-D fast imaging using a single surface coil for pulse transmission and signal detection. Computer calculations and phantom, rat, and human surface coil imaging experiments demonstrate the utility of this adiabatic pulse for T1-weighted imaging with a transmitter coil which generates a highly-inhomogeneous B1 field profile.
| Original language | English (US) |
|---|---|
| Pages (from-to) | 559-567 |
| Number of pages | 9 |
| Journal | Investigative Radiology |
| Volume | 25 |
| Issue number | 5 |
| DOIs | |
| State | Published - May 1990 |
Keywords
- Adiabatic pulses
- Flash
- Mri
- Surface coil
- T
- relaxation
Fingerprint
Dive into the research topics of '3-D FLASH Imaging using a single surface coil and a new adiabatic pulse, BIR-4'. Together they form a unique fingerprint.Cite this